Design Verification Urea liquicolor Contents 1 Introduction .................................................................................................................................................................. 2 2 Imprecision .................................................................................................................................................................... 2 3 Linearity and Detection Limit ...................................................................................................................................... 2 Linearity......................................................................................................................................................................................................... 2 Detection Limit ........................................................................................................................................................................................... 3 4 Recovery of Control Sera .............................................................................................................................................. 3 5 Comparison of Methods ............................................................................................................................................... 4 6 Interferences ................................................................................................................................................................. 6 7 Stability.......................................................................................................................................................................... 8 Recovery in Control Sera ......................................................................................................................................................................... 8 Linearity......................................................................................................................................................................................................... 8 Temperature Stress Results ................................................................................................................................................................... 8 8 Stability Working Reagent ........................................................................................................................................... 9 Recovery in Control Sera ......................................................................................................................................................................... 9 Linearity......................................................................................................................................................................................................... 9 9 Traceability .................................................................................................................................................................... 9 Form: 4.2-01.06-04 Rev. 009 | valid of 25.11.2020 1/9 1 Introduction The imprecision and the recovery of the UREA liquicolor test has been tested and documented in order to verify its clinical usefulness and additional features such as linearity, interferences and method comparison. 2 Imprecision The imprecision (within-run and day-to-day) of the UREA liquicolor method was calculated from six determinations on six consecutive days. Three pooled control sera were employed as sample material. Analyser: HITACHI 717 Analyte concentration (mg/dl) 27.3 63.3 276 Intra-assay SD (mg/dl) 1.042 1.464 3.51 Inter-assay %CV 3.82 2.31 1.27 SD (mg/dl) 1.115 2.557 11.157 %CV 4.08 4.04 4.04 3 Linearity and Detection Limit Linearity The linearity of the UREA liquicolor method was controlled by employing dilutions of a patient serum pool which has been spiked with urea to approx. 450 mg/dl. The analysed concentrations were calculated vs. the regression line. The deviation between analytical and regressed data are expressed in absolute and relative values. As an example data measured on the Hitachi 717 are reported. High Pool Content (%) 0 10 20 30 40 50 60 70 80 90 100 Analytical Data mg/dl 1.70 43.7 84.7 124 168 216 253 303 338 380 420 Design Verification and Product Data for UREA LIQUICOLOR Rev. 009 Regressed Data mg/dl 0.14 42.7 85.2 128 170 213 255 298 340 383 426 Deviation from Regression Line mg/dl (%) 1.56 1076.92 1.01 2.38 -0.53 -0.62 -3.62 -2.83 -1.91 -1.12 2.70 1.27 -1.89 -0.74 5.06 1.70 -2.38 -0.70 -3.22 -0.84 -5.56 -1.31 2/9 LINEARITY 35 Linear Regression Error Bars = ± 2SD Measured Analyte Concentration (mg/dl) 30 25 20 15 10 5 0 0 20 40 60 80 100 120 High Pool Content (%) Conclusion: The UREA liquicolor test is linear up to an urea concentration of at least 400 mg/dl. Detection Limit A 20-fold determination of a ‘0’ sample (phys. saline) on a Hitachi 717 analyser revealed an absolute mean of -0.65mg/dl and a SD of 1.47mg/dl. From the three-fold standard deviation the detection limit can therefore be calculated on the base of mean +3 SD to 5.06mg/dl. 4 Recovery of Control Sera A number of commercially available control sera have been employed. The control sera have been reconstituted/prepared according to the manufacturer’s instructions. Five-fold determinations of each control serum have been performed with UREA liquicolor reagents of different production lots. The mean values have been calculated and compared with the target values of the respective control sera. CONTROL SERUM RECOVERY AT 578 nm Control LOT serum Fresh Reference 10 days 37°C 12 days 37°C Target Range Result Deviation Result Deviation Result Deviation Result Deviation mg/dl mg/dl mg/dl (%) mg/dl (%) mg/dl (%) mg/dl (%) -8.14 HumaTrol N # 018 51.1 38.9 – 63.5 44.2 -13.46 41.3 -19.22 45.5 -10.96 46.9 HumaTrol P # 016 172.0 131 – 213 157.2 -8.56 149.4 -13.12 469.3 -1.57 174.5 1.48 SERODOS # 6868 52. 39.5 – 64.5 54.1 4.08 52.1 0.23 57.3 10.27 59.3 14.08 SERODOS # 6795 172.7 131.3 – 157.3 -8.91 152.8 -11.52 166.7 -3.50 174.8 1.22 plus 214.1 Precinorm # 154916 43.2 32.7 – 53.7 45.3 4.91 43.3 0.19 43.9 1.71 45.1 4.49 Precipath # 199459 140 107 - 173 152.5 8.91 146.4 4.60 150.7 7.64 155.6 11.16 Summary Conclusion: 610.7 585.36 633.44 656.4 No significant differences could be observed between fresh and stressed reagents, confirming the stability of the reagents. Controls were all recovered well within the specified ranges. As an alternative measurement wavelength, control recovery of fresh material was also assayed at 546nm. The mean values of two determinations each were calculated and compared to the reference range of the respective control sera. The corresponding recoveries at 578 nm were also measured for comparison. Design Verification and Product Data for UREA LIQUICOLOR Rev. 009 3/9 CONTROL SERUM RECOVERY Control LOT serum Fresh at 546 nm Fresh at 578 nm Target Range Result Deviation Result Deviation mg/dl mg/dl mg/dl (%) mg/dl (%) HumaTrol N # 001 97.2 75.8 – 119.0 85.3 - 12.2 87.3 - 10.2 HumaTrol P # 001 281.0 219 – 343 262.8 - 6.5 276.1 - 1.7 SERODOS # 001 88.3 68.9 – 108.0 81.5 - 7.7 80.7 - 8.6 SERODOS plus # 6798 143.0 112 - 174 140.1 - 2.0 143.6 0.4 Precinorm # 150403 43.8 37.2 – 50.4 39.3 - 10.3 39.0 - 11.0 Precipath # 150410 150 126- 174 139.9 - 6.7 143.1 - 4.6 748.9 - 6.8 769.8 - 4.2 Sum 803.3 Conclusion: All control sera were found within the respective reference ranges. Therefore, measurement at 546 nm is feasible. 5 Comparison of Methods The UREA liquicolor test has been compared against a commercially available urea method. Control sera as well as patient samples (serum and urine) have been employed in the comparison. The results have been evaluated by a main component analysis. The linear regression obtained could be described as follows: Serum: N= r= Y= Xmean = Ymean = Sample Reference no. 1 UREA liquiUV 1 27.1 2 24.9 3 52.5 4 22.8 5 35.9 6 38.2 7 48.5 8 53.1 9 25.4 10 41.8 11 17.8 12 21.0 13 25.6 14 28.3 15 53.6 16 35.9 17 40.0 18 31.0 19 44.5 20 23.8 21 25.0 22 37.2 23 25.7 24 38.4 25 33.6 26 40.1 27 28.9 28 23.1 29 32.9 30 23.0 31 42.6 32 55.5 33 53.8 Test 2 26.5 24.0 53.2 23.3 36.1 39.2 48.6 52.9 25.4 43.1 18.5 21.0 26.4 26.2 52.9 35.9 39.4 30.0 44.1 23.0 26.4 35.7 25.4 40.1 36.0 41.7 29.0 22.2 33.2 23.0 40.2 54.3 52.9 1 UREA liquicolor 27.5 24.8 52.1 23.1 35.9 37.7 46.1 51.4 25.1 41.2 17.8 20.9 25.1 26.7 51.8 34.6 37.9 29.3 44.3 22.8 25.1 35.0 25.6 39.0 34.6 39.4 28.1 22.4 32.7 23.8 42.1 56.2 54.2 Design Verification and Product Data for UREA LIQUICOLOR Rev. 009 55 0.998 0.981 X + 0.405 35.53 35.27 Mean Reference 2 28.0 24.5 53.5 23.3 36.4 38.0 47.4 50.8 25.1 41.5 17.7 21.2 25.1 26.1 51.6 34.2 38.9 28.8 44.5 22.9 24.5 35.8 24.3 38.6 35.2 40.2 28.8 22.2 32.4 23.5 41.8 54.9 53.8 26.8 24.45 52.85 23.05 36 38.7 48.55 53 25.4 42.45 18.15 21 26 27.25 53.25 35.9 39.7 30.5 44.3 23.4 25.7 36.45 25.55 39.25 34.8 40.9 28.95 22.65 33.05 23 41.4 54.9 53.35 Test 27.75 24.65 52.8 23.2 36.15 37.85 46.75 51.1 25.1 41.35 17.75 21.05 25.1 26.4 51.7 34.4 38.4 29.05 44.4 22.85 24.8 35.4 24.95 38.8 34.9 39.8 28.45 22.3 32.55 23.65 41.95 55.55 54 Regression 26.70 24.40 52.26 23.02 35.73 38.38 48.04 52.41 25.33 42.06 18.21 21.01 25.92 27.14 52.66 35.63 39.36 30.33 43.87 23.37 25.62 36.17 25.48 38.92 34.55 40.54 28.81 22.63 32.84 22.97 41.03 54.28 52.75 Deviation 0.95 0.2 -0.05 0.15 0.15 -0.85 -1.8 -1.9 -0.3 -1.1 -0.4 0.05 -0.9 -0.85 -1.55 -1.5 -1.3 -1.45 0.1 -0.55 -0.9 -1.05 -0.6 -0.45 0.1 -1.1 -0.5 -0.35 -0.5 0.65 0.55 0.65 0.65 4/9 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 24.9 44.5 23.5 20.9 72.2 42.0 72.5 46.1 28.1 23.1 22.4 24.6 27.5 33.0 24.1 31.2 37.2 51.8 34.1 33.7 36.9 52.3 24.4 42.2 22.7 20.9 71.4 42.2 71.4 44.2 26.7 22.8 22.9 23.5 27.7 35.8 23.4 32.0 34.8 52.3 35.6 33.4 36.5 53.2 25.4 44.6 24.1 22.0 70.7 43.4 70.3 45.1 27.1 23.2 23.0 23.6 28.2 33.8 23.2 31.4 38.0 51.8 33.5 32.1 37.8 52.1 25.7 44.4 24.0 21.6 70.5 42.7 71.9 44.5 26.9 22.6 23.6 23.8 28.4 34.2 24.4 30.7 38.7 51.7 34.0 32.2 37.4 51.1 24.65 43.35 23.1 20.9 71.8 42.1 71.95 45.15 27.4 22.95 22.65 24.05 27.6 34.4 23.75 31.6 36 52.05 34.85 33.55 36.7 52.75 25.55 44.5 24.05 21.8 70.6 43.05 71.1 44.8 27 22.9 23.3 23.7 28.3 34 23.8 31.05 38.35 51.75 33.75 32.15 37.6 51.6 24.59 42.94 23.07 20.91 70.86 41.72 71.01 44.71 27.29 22.92 22.63 24.00 27.49 34.16 23.71 31.41 35.73 51.48 34.60 33.33 36.42 52.17 0.9 1.15 0.95 0.9 -1.2 0.95 -0.85 -0.35 -0.4 -0.05 0.65 -0.35 0.7 -0.4 0.05 -0.55 2.35 -0.3 -1.1 -1.4 0.9 -1.15 UREA liquicolor Serum 80 70 Test mg/dl 60 50 40 30 20 10 10 20 30 40 50 60 70 80 Reference mg/dl Urine: N= r= Y= Xmean = Ymean = Sample Reference no. 1 UREA liquiUV 1 2091 2 2889 3 818 4 1626 5 1131 6 1212 7 1131 8 1980 9 2454 10 2454 11 1242 12 1919 13 1768 14 970 15 616 Test 2 2050 2838 818 1646 1293 1242 1141 2000 2464 2424 1252 1919 1727 1000 586 1 UREA liquicolor 2104 2761 903 1630 1347 1286 1165 1943 2337 2387 1286 1943 1741 963 670 Design Verification and Product Data for UREA LIQUICOLOR Rev. 009 38 0.995 0.974 X + 12.428 1451.08 1425.84 Mean Reference 2 2054 2842 812 1610 1317 1175 1195 1842 2549 2448 1195 1781 1700 903 660 2071 2863 818 1636 1212 1227 1136 1990 2459 2439 1247 1919 1747 985 601 Test 2079 2801 857 1620 1332 1231 1180 1892 2443 2418 1241 1862 1721 933 665 Regression 2029 2801 809 1606 1193 1208 1119 1950 2408 2388 1227 1882 1714 972 598 Deviation 8.7 -62.0 38.9 -16.6 119.8 3.6 44.0 -97.4 -16.6 -21.7 -6.5 -57.0 -26.7 -52.0 64.2 5/9 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 2404 1151 657 2454 394 1242 1707 2596 1525 1697 475 1495 949 606 1151 1172 1576 586 1505 1465 2212 1434 333 2222 1121 586 2414 404 1273 1737 2687 1545 1656 535 1525 899 566 1162 1202 1707 596 1485 1414 2343 1374 343 2428 1165 600 2438 509 1246 1620 2690 1468 1711 398 1256 802 680 1125 1246 1660 660 1347 1367 2185 1286 317 2306 1165 600 2357 468 1226 1539 2529 1418 1620 398 1438 923 509 1054 1155 1650 529 1296 1549 2165 1347 377 2313 1136 621 2434 399 1257 1722 2641 1535 1677 505 1510 924 586 1156 1187 1641 591 1495 1439 2278 1404 338 2367 1165 600 2397 488 1236 1579 2609 1443 1665 398 1347 862 594 1089 1200 1655 594 1322 1458 2175 1317 347 2265 1119 617 2383 401 1237 1690 2585 1508 1646 504 1483 913 583 1139 1168 1611 588 1468 1414 2231 1380 342 54.1 28.8 -21.7 -36.8 89.5 -21.6 -142.9 -31.8 -92.3 -11.6 -107.5 -163.1 -62.1 8.7 -67.1 13.7 13.7 3.6 -173.2 18.8 -102.5 -87.3 8.6 U R E A liq u ic o lo r U rin e 3000 2500 Test mg/dl 2000 1500 1000 500 0 0 500 1000 1500 2000 2500 3000 R e fe re n c e m g / d l Conclusion: Both methods showed a good agreement and no significant deviation could be observed with any specific sample. 6 Interferences Interference by triglycerides, hemoglobin, bilirubin and ascorbic acid at 578 nm have been studied by adding known amounts of the potentially interfering substance to a known sample. Recoveries have been analyzed according to the method of Glick et al. (Clin.Chem. 1986, 32 470-5). The results are summarised in the following table. Triglyc. Analyt. Dev. Hemogl Result Analyt. Dev. Bilirubi Result Analyt. Dev. Result Ascob. Analyt. acid Result Dev. mg/dl mg/dl % mg/dl mg/dl % mg/ml mg/dl % mg/ml mg/dl % 0 55.60 100 0 64.20 100 0 64.80 100 0 66.95 100 250 58.65 105.5 50 64.45 100.4 4 64.95 100.2 2 66.80 99.8 500 61.45 110.5 100 63.85 99.5 8 64.90 100.2 4 66.95 100.0 750 62.75 112.9 150 63.70 99.2 12 64.95 100.2 6 66.75 99.7 1000 64.00 115.1 200 64.75 100.9 16 63.80 98.5 8 67.10 100.2 1250 66.80 120.1 250 64.05 99.8 20 64.40 99.4 10 67.25 100.4 1500 69.05 124.2 300 64.00 99.7 24 64.65 99.8 12 67.40 100.7 1750 70.70 127.2 350 63.45 98.8 28 64.20 99.1 14 66.60 99.5 2000 72.65 130.7 400 63.00 98.1 32 64.85 100.1 16 66.55 99.4 Design Verification and Product Data for UREA LIQUICOLOR Rev. 009 6/9 2250 73.70 132.6 450 64.75 100.9 36 64.30 99.2 18 67.80 101.3 2500 76.20 137.1 500 64.50 100.5 40 64.35 99.3 20 66.95 100.0 Glick 3 1 1 1 Conclusion: No significant interfering effect has been seen up to following concentrations: Hemoglobin 500 mg/dl Bilirubin 40 mg/dl Ascorbate 20 mg/dl. Triglycerides showed a positive interfering effect of about 10% at 500 mg/dl and about 15% at 1000 mg/dl triglycerides. Furthermore, interferences of haemoglobin and bilirubin at 546 nm measurement wavelength have been studied by adding known amounts of the potentially interfering substance to a known sample. Recoveries have been analyzed according to the method of Glick et al. (Clin.Chem. 1986, 32 470-5). Hemogl Analyt. Dev. Bilirubi Result mg/dl mg/dl Analyt. Dev. Result mg/dl % mg/ml 0 59.40 100 0 61.10 100 % 50 55.25 93.0 4 60.60 99.2 100 52.85 89.0 8 60.90 99.7 150 50.70 85.4 12 59.75 97.8 200 49.35 83.1 16 59.35 97.1 250 48.55 81.7 20 59.85 98.0 300 46.15 77.7 24 59.55 97.5 350 45.25 76.2 28 60.35 98.8 400 45.05 75.8 32 61.10 100.0 450 44.10 74.2 36 60.45 98.9 500 43.25 72.8 40 60.15 98.4 Glick 3 1 Conclusion: No significant interfering effect has been seen up to following concentrations: Bilirubin 40 mg/dl Hemoglobin interferes significantly (>10% deviation) above 50 mg/dl. Design Verification and Product Data for UREA LIQUICOLOR Rev. 009 7/9 7 Stability Recovery in Control Sera For real-time study, one of the available multipurpose reagent kit sizes was used. The UREA liquicolor multipurpose reagent was stored at 2…8°C. The real-time stability was tested up to 125% of the shelf life at several intervals. CONTROL SERUM RECOVERY Control LOT serum Reference 23 months 24 months 26 months Target Range Result Deviation Result Deviation Result Deviation Result Deviation mg/dl mg/dl mg/dl (%) mg/dl (%) mg/dl (%) mg/dl (%) HumaTrol N # 018 51.2 38.9 – 63.5 40.8 1.25 41.5 1.23 41.7 1.22 41.1 1.24 HumaTrol P # 016 172 131 - 213 150.8 1.13 151.6 1.13 41.6 1.22 150.8 1.14 SERODOS # 6868 52 39.5 – 64.5 50.6 1.02 51.7 1.00 50.8 1.02 50.6 1.02 SERO.Plus # 6795 173 131 - 214 151.2 1.13 154.4 1.12 153.3 1.12 151.3 1.14 Precinorm #159296 42.2 32.0 – 52.4 44.8 0.95 44.0 0.96 43.9 0.96 43.4 0.97 Precipath #165215 146 110 - 182 146 110 - 182 147.3 0.99 146.9 0.99 144.4 1.00 Mean 106 97.3 98.4 97.7 96.9 Conclusion: The control recovery data for UREA liquicolor multipurpose reagent are indicative for a real-time stability of 24 months at 2…8°C. Linearity Four independent lots have been stored under the recommended conditions for up to 26 months 2…8°C. A high concentration pool serum was employed and dilutions with phys. saline were made. The analysed concentrations were calculated vs. the regression line. Deviation from the regression line are expressed in absolute and relative values. Reference 20 months 24 months 26 months LOT H059 R1 LOT CD321 R1 LOT UJ 074 LOT H011 R2 LOT CD622 R2 LOT UJ060 Enz. LOT CD620A Enz. LOT UJ061 High pool Analytical Data Regression Analytical Data Regression Analytical Data Regression Analytical Data Regression Content (%) mg/dl Line (%) mg/dl Line (%) mg/dl Line (%) mg/dl Line (%) 0 -0.70 -137.23 1.15 -373.81 0.45 -314.29 1.00 -166.67 10 54.20 3.95 42.20 -2.62 44.95 3.21 52.50 14.01 20 104 1.56 87.09 -0.85 84.65 -3.05 76.30 -18.48 30 154 0.62 131 -1.07 130 -1.16 146 3.72 40 201 -1.00 175 0.97 177 1.21 192 1.64 50 253 -0.11 218 0.64 221 1.15 241 2.14 60 301 -0.64 262 0.42 265 1.09 293 3.17 70 347 -1.98 306 0.83 311 1.48 347 4.63 80 387 -4.15 350 1.08 351 0.47 388 2.48 90 421 -7.23 393 2.59 401 1.83 434 1.70 100 460 -8.90 437 -0.04 435 -0.44 472 -0.42 Temperature Stress Results In addition to the real-time storage, temperature stress tests have been applied on the UREA liquicolor test. Stressed reagents have been compared with fresh ones (see chapter 4). Conclusion: The results obtained with reagents stored for even 26 months under real-time conditions and with stressed reagents. From the above linearity comparison the stability claim of 24 months 2…8°C from the date of manufacture has been confirmed. Design Verification and Product Data for UREA LIQUICOLOR Rev. 009 8/9 8 Stability Working Reagent Recovery in Control Sera The working reagent has been stored 3 weeks at 25°C and 4 weeks at 2…8°C. Commercially available control sera have been employed. The control sera have been reconstituted/prepared according to the manufacturer’s instructions. Twofold determinations of each control serum have been performed with the UREA liquicolor test method. The means of the twofold determinations have been calculated and compared with the target values. CONTROL SERUM RECOVERY Control LOT serum 3 weeks 25°C 4 weeks 2…8°C Target Range Result Deviation Result Deviation mg/dl mg/dl mg/dl (%) mg/dl (%) HumaTrol N # 016 51.1 38.9-63.5 45.9 -10.22 49.2 -3.76 HumaTrol P # 018 172 131-213 165.7 -3.65 165.7 -3.69 SERODOS # 6868 52 39.5-64.5 58.4 12.38 53.7 3.27 SERODOS plus # 6795 172.7 131.3-214.1 171.5 -0.72 160.8 -6.88 Precinorm #154916 43.2 32.7-53.7 45.5 5.32 44.1 1.99 Precipath #199459 140 107-173 154.3 10.20 150.5 7.47 641.28 13.32 623.9 -1.59 Summary 1074 Linearity The working reagent has been stored 3 weeks at 25°C and 4 weeks at 2…8°C. A high concentration pool serum was employed and dilutions with phys. saline were made. The analysed concentrations were calculated vs. the regression line. Deviation from the regression line are expressed in absolute and relative values. Reference 3 weeks 25°C 4 weeks 2…8°C High pool Analytical Data Regression Line Analytical Data Regression Line Analytical Data Content (%) mg/dl (%) mg/dl (%) mg/dl (%) 0 1.70 1077 0.00 -100 0.00 -100 10 43.7 2.38 41.6 0.37 40.5 -1.38 20 84.7 -0.62 86.3 -2.45 84.5 -0.54 30 124 -2.83 132 -2.50 128 -0.58 40 168 -1.12 178 -2.27 170 -1.64 50 216 1.27 229 -0.33 217 0.29 60 254 -0.74 278 0.40 260 -0.31 70 303 1.70 325 0.29 303 -0.49 80 338 -0.70 373 0.72 352 0.98 90 380 -0.84 424 1.62 393 0.27 100 420 -1.31 461 -0.74 430 -1.45 Conclusion: Regression Line The above results clearly demonstrate the stability of the working reagent for 2 weeks at 25°C and for 4 weeks at 2…8°C. 9 Traceability The UREA liquicolor test has to be calibrated with a standard. UREA concentration in the kit standard has been verified with the reference material NIST SRM 909c. UREA liquicolor revealed recovery well within the reproducibility limits of the method. Conclusion: The method is traceable to the reference preparation NIST SRM 909c. Design Verification and Product Data for UREA LIQUICOLOR Rev. 009 9/9