STA Satellite Max
ASTM Protocol
Table of contents
1
Reference Standards ........................................................................................... 5
2
Scope.................................................................................................................... 5
3
Limitation of the document ................................................................................. 5
4
Hardware Interface .............................................................................................. 5
5
Definitions ............................................................................................................ 6
6
Standard exchanges types ................................................................................. 6
7
Instrument Setup ................................................................................................. 7
7.1
LIS Connection Setup ......................................................................................................7
8
Detailed Message and Records Description ..................................................... 9
8.1
Message Header Record .................................................................................................9
8.2
Patient Information Record ............................................................................................10
8.3
Test Order Record .........................................................................................................11
8.4
Result Record ................................................................................................................12
8.5
Comment Record...........................................................................................................14
8.6
Request Information Record ..........................................................................................15
8.7
Message Terminator Record..........................................................................................16
8.8
Scientific Record .................................................................... ...........................................17
8.9
Manufacturer Information Record...................................................................................17
8.10
Error and Flag Code Table.............................................................................................18
9
Examples of Messages sent to the LIS ............................................................ 19
9.1
Queries ..........................................................................................................................19
9.1.1
Worklist Query ................................................................................................................................ 19
9.1.2
Worklist Query Cancelled ............................................................................................................... 19
9.2
Results Upload ..............................................................................................................19
9.2.1
Patient Result Upload ..................................................................................................................... 19
9.2.2
Patient Result Upload with Alarms and Patient Information ........................................................... 20
9.2.3
Patient Result Upload – Result in Duplicate ................................................................................... 20
9.2.4
Control Result Upload .................................................................................................................... 21
0932500_EN
xxxx
3/21
20
STA Satellite Max
ASTM Protocol
Revision table
Reference
0932500_EN
4/21
Date
March 2018
Description
Creation
0932500_EN
STA Satellite Max
ASTM Protocol
1 Reference Standards
The ASTM connection protocol on the STA Satellite Max is based on the two following standards:

CLSI LIS01-A2 : Specification for Low Level protocol to Transfer Messages Between Clinical
Laboratory Instruments and Computer System; Approved Standard - Second Edition (Previously
ASTM E1381-95)

CLSI LIS2-A2 : Specification for Transferring Information Between Clinical Laboratory
Instruments and Information Systems; Approved Standard - second Edition (Previously
ASTM E1394-97)
2 Scope
This Document applies to STA Satellite Max GUI software version 1.0.18.2 and above.
3 Limitation of the document
This document does not describe the ASTM communication protocol. For details about the low level
and high level communication protocol details please refer the Standards referenced in section 1.
Reference Standards
4 Hardware Interface
The LIS communication protocol is implemented on a serial RS232 physical transport link, with
following characteristics:
Parameter
Interface
Value
Serial type, standard RS232
Or via
USB to serial adapter
SUBD 9 - Male on the instrument (PC) side
Pin 2 : Rx (Reception)
Connector and pin-out
Pin 3 : Tx (Transmission)
Pin 5 : Ground
Flow Control / Handshaking
Pin 1, 4, 6, 7, 8, 9 : Not connected
None
Baud Rate*
110 to 115200 Bauds (19200 by default)
Parity*
Even, Odd, Mark, Space, None (Odd by default)
Data Bit*
5, 6, 7, 8 (8 by default)
Stop Bit*
1,2 (1 by default)
* These parameters have to be specified in the [Software Configuration] ; [LIS settings] tab of
the analyzer.
estem
0932500_EN
xxxx
5/21
20
STA Satellite Max
ASTM Protocol
5 Definitions
Term
Definition
ASTM
American Association for Test and Material
L.I.S / LIS
Laboratory Information System
N.T.
Not Transmitted
Text
A string of characters in range permitted by the Standard
Date-Time
A string of character defining the Day, Month, Year, Hour
Minute and second.
N/A
Not applicable
NULL
An empty string of characters
Download (D)
Transmission of data from the LIS to the Instrument
Upload (U)
Transmission of data from the Instrument to the LIS
6 Standard exchanges types
The following exchanges are supported by the STA Satellite Max
#
1
Direction
Instrument  LIS (U)
Description
Worklist Query
Pattern
H.Q.L
H.P.O.[R.[C.].M]L
2
Instrument  LIS (U)
3
LIS  Instrument (D)
Worklist
download
H.P.O.L
4
LIS  Instrument (D)
Result Query
H.Q.L
Results Upload
([] can be repeated, C
& M not Mandatory)
With:
6/21
H:
Header Record
P:
Patient Record
O:
Order Record
Q:
Query Record
R:
Result Record
C:
Comment Record
M:
Manufacturer Record
L:
Terminator Record
0932500_EN
Description
The instrument sends this query to
the LIS to request the Worklist of a
given sample. This is done
automatically at loading time, if the
option has been selected.
The instrument uses this message to
send back results to the LIS. It can
be done automatically when the
result is available if the option has
been selected (see section 7.1).
The results can also be sent to the
LIS on request from the user from the
This
message
is as
used
by the LIS
archive
menu or
a response
to to
the
send
the workload
of a4)given sample
LIS result
query. (See
to the instrument. It can be done as
a response to the Worklist query
(see 1) or as a test or rerun request
from the LIS. It can also be used to
change test priority.
This message is used by the LIS
to get test results from the
Instrument.
STA Satellite Max
ASTM Protocol
7 Instrument Setup
7.1
LIS Connection Setup
LIS serial configuration:
Term
Definition
Baud Rate
Define the communication speed in Bauds (bit/s)
Data Bits
Define the number of data bits used to code each character
Stop Bits
Defines the number of stop bits
Parity
Define the type of parity check used
Increment Specimen Sequence
Define how the Instrument Specimen ID field will be
generated by the STA Satellite Max . (See section 8.3 for a
description)
Control Transmission enabled
If YES is checked, control results will be transmitted to the
LIS
Auto Transfer Enabled
If YES is checked, the results will be automatically
transmitted to the LIS.
Host Query Enabled
If YES is checked, the instrument will query automatically
the LIS for the Worklist when loading a sample.
Sample Re-query allowed
If YES is checked, the instrument will query automatically
the LIS for the Worklist each time the sample is loaded.
Otherwise the query will be sent only when the sample is
loaded the first time.
Note: This field will be displayed only if the ‘Host Query’ option is enabled.
Sample Re-query allowed
The time The STA Satellite Max will wait for a response
after a Sample Worklist Query. After this delay, the query is
cancelled. ( see sections 8.6 and 9.1).
Button to test communication setting with the LIS.
0932500_EN
xxxx
7/21
20
STA Satellite Max
ASTM Protocol
1
Test Setup :
Term
8/21
Definition
LIS Code
Test ID sent to the LIS
Internal code
Internal code of the test sent to the LIS
0932500_EN
STA Satellite Max
ASTM Protocol
8 Detailed Message and Records Description
The Section references are given for the CLSI document. The former ASTM document numbering is
given between brackets.
8.1
Message Header Record
CLSI LIS2-A2
(ASTM 1394)
Value
Field Description
Format
Record Type ID
1 ASCII character
U D
Comment
Section
6.1 (7.1.1)
‘H’
6.2 (7.1.2)

Delimiter
Definition
‘|’
‘\’
‘^’
‘&’
6.3 (7.1.3)
‘|’
6.4 (7.1.4)
‘|’
Field Delimiter
Repeat Field
delimiter
Component
DelimiterDelimiter
Escape
1 ASCII
1 ASCII
characters
1
ASCII
characters
1
ASCII
characters
characters

Message control /
ID
/
Access
Not used
Not used
password
6.5 (7.1.5)
Sender Name or
ID
‘STA_Satellite_max’
‘STA_Satellite_max’
The Serial Number is stored in the
Instrument CPU board memory. The
Version Number identifies the
implementation of the protocol, not of the
software. The Version Number cannot be
changed
4 ASCII
characters
Name
‘
‘^’

Serial Number
10 ASCII
characters
‘1’
Version Number
1 ASCII
characters
6.6 (7.1.6)
‘|’
Sender mailing
address
/

Not used
6.7 (7.1.7)
‘|’
Reserved field
/

Not used
6.8 (7.1.8)
‘|’
Sender
telephone
number
/

Not used
6.9 (7.1.9)
‘|’
Sender
characteristics
/

Not used
Receiver ID
N/A
checked by the STA Satellite Max
Comments or
special
instructions.
/
Text
‘^’
6.10 (7.1.10)
6.11 (7.1.11)
Ignored
‘|’
1 ASCII character
6.12 (7.1.12)
‘P’ or ‘D’
6.13 (7.1.13)
‘1’
6.14 (7.1.14)
Date-Time
‘P’ : Production
(normal process)
‘D’ : Debug
(Ignored)
Processing ID
Version Number
1 ASCII character
The Receiver ID is accepted but is not

Only the processing ID ‘P’ is accepted and
transmitted by the STA Satellite Max. All

 other Processing IDs are ignored (D) and
not transmitted (U).

Identifies the version level of the
specification.
In download mode, this field is mandatory.
14 ASCII
The message will not be recognized by the
Date and Time of characters
 STA Satellite Max if this field is not present.
Message
‘YYYYMMDDHHM
MSS’
0932500_EN
xxxx
9/21
20
STA Satellite Max
8.2
ASTM Protocol
Patient Information Record
CLSI LIS2-A2
(ASTM 1394)
Value
Field Description
Format
U D
Comment
Section
7.1 (8.1.1)
‘P’
7.2 (8.1.2)
Record Type ID
1 ASCII character
 
Sequence Number
N ASCII characters
  Field size not limited
/

Up to 255 ASCII
characters
 
/

Not used
/

Not used
>1
7.3 (8.1.3)
7.4 (8.1.4)
7.5 (8.1.5)
‘|’
Patient ID
‘|’
Practice
Assigned
Patient ID
Laboratory
Assigned
Patient ID
Patient ID No 3
Not used
Item1
^
Item2
^
Item3
^
Item4
Both the Name and the First Name must be
Patient Name
7.6 (8.1.6)
Text
^
Text
7.8 (8.1.8)
Date
7.9 (8.1.9)
Ignored
 specified for the Patient Name field to be taken
Last Name
255 ASCII characters
max
First Name
255 ASCII characters
max
Birth Date
ASCII ‘YYYYMMDD’
 
Patient Sex
1 ASCII character

‘M’:
’F’:
’U’:
7.19 (8.1.19)
10/21
The Patient Sex field is accepted but ignored by
the STA Satellite Max. It is never sent by the STA
 Satellite Max
Both the Name and the First Name must be
 specified for the Patient Name field to be taken
Text
^
Last Name
255 ASCII characters
max
Text
First Name
255 ASCII characters
max
Patient’s known or
suspected
diagnosis
N/A
Ignored
In Result Upload mode, If the specimen is a
Control the Last name will be the Assay Internal
Code as defined in the STA Satellite Max
Application (See 7.2) and the first Name will be
L_n where n is the control level
Male
Female
Unknown
Patient Name
7.14 (8.1.14)
into account.
0932500_EN
into account.
In Result Upload mode, If the specimen is a
Control the Last name will be the Assay Internal
Code as defined in the STA Satellite Max
Application (See 7.2) and the first Name will be
L_n where n is the control level

The Diagnosis field is accepted but ignored by
STA Satellite Max. It is never sent by the STA
Satellite Max
STA Satellite Max
ASTM Protocol
CLSI LIS2-A2
(ASTM 1394)
Value
Field Description
Format
U D
Comment
Section
7.20 (8.1.20)
Ignored
Patient’s active
Medications
N/A

The Medication field is accepted but ignored by
the STA Satellite Max. It is never sent by the
STA Satellite Max
7.25 (8.1.25)
Ignored
Admission Status
N/A

The Status field is accepted but ignored by the
STA Satellite Max. It is never sent by the STA
Satellite Max
7.26 (8.1.26)
Text
Location
255 ASCII characters
max
 
7.34 (8.1.34)
Ignored
Hospital Institution
N/A

8.3
The Institution field is accepted but ignored by
the STA Satellite Max. It is never sent by the
STA Satellite Max
Test Order Record
CLSI LIS2-A2
(ASTM 1394)
Format
U
D
Value
Field Description
Comment
8.4.1 (9.4.1)
‘O’
Record Type ID
1 ASCII character
 
8.4.2 (9.4.2)
>1
Sequence
Number
N ASCII characters
  Field size not limited
Section
8.4.3 (9.4.3)
Patient ID
Text
^
Barcode ID
255 ASCII characters max
Ignored / N.T
^
Accession ID
Ignored
Text
^
Carrousel Barcode
255 ASCII characters max
Carrousel Position
2 ASCII characters
Instrument
Specimen ID
1 to 9999
Text
8.4.4 (9.4.4)
  Accession ID is accepted but ignored by the
Specimen ID
Id (Text)
STA Satellite Max. It is never transmitted by the
STA Satellite Max.
In Result Upload mode, If the specimen is a
Control the Barcode Id will be : TIC CL (Test
Internal Code – Control Level)
Where TIC is the Assay Internal Code as
defined in the STA Satellite Max.
Application (See 7.2) and CL is L_n where n is
the control level.

0932500_EN
xxxx
This field is generated by the STA Satellite Max
as a number between 1 and 9999.
The STA Satellite Max will affect either the same
Instrument specimen ID to all the results of a
given specimen or will generate a new specimen
ID for each result depending on the flag
‘Increment Specimen Sequence’ described in
section 7.1.
The current value of the Instrument Specimen ID
is stored in the Pc and the counting will continue
in sequence at each software launch.
11/21
20
STA Satellite Max
CLSI LIS2-A2
(ASTM 1394)
Value
ASTM Protocol
Field Description
U
Format
D
Comment
Section
Universal Test ID
8.4.5 (9.4.5)
Component 1-3 are
NULL
^
^
^
Test Code (Text)
^
‘#’,’F’
The value of the Test Code (component 4) is
user defined in the Assay Definition. (See 7.2)
Assay Code
5 ASCII characters max
Result Type
^
1 ASCII character
‘F’: Final Result
‘#’ : ‘1’ or ‘2’ for duplicates
‘48’ or ‘64’
^
Plasma Type
‘48’ : Control
‘64’ : Specimen
‘90’, ‘91’ or ‘92’
^
Component Name
2 ASCII characters
‘90’ : Control Level 1
‘91’ : Control Level 2
‘92’ : Control Level 3
Dilution (Text)
^
Dilution
‘X:Y’ NULL if no Dilution
Ignored
^
Reagent Lot Number N/A
Ignored
Plasma Lot Number
Priority
8.4.6 (9.4.6)
8.4.8 (9.4.8)
‘S’, ‘R’
‘S’ : Stat
‘R’ : Routine
Specimen Collection
date and Time
Action Code
‘A’, ‘N’, ‘C’
 ‘A’ and ‘S’ are both interpreted as Stat

N/A
 
1 ASCII character
‘A’ : Add Requests
‘N’ : New Requests
‘C’ : Cancel Requests
Report Type
8.4.26 (9.4.26)
The F flag (Final result) is used to notify that a
result has been obtained either by averaging two
 duplicate results or by applying a mathematical
formula defined for a calculated test. The
different results that are used to compute the
final results are flagged with ‘1’ and ‘2’
(duplicates).
1 ASCII character
‘S’, ‘A’ : Stat
Ignored / N.T.
STA Satellite Max Universal Test ID max length
is 5
N/A
‘S’, ‘A’
8.4.12 (9.4.12)
X’,‘F’,‘Z’,’Q’
‘O’,’F’,’Y’,’Z’,’Q’
8.4
This field is NULL in Upload mode ( Result
transmission from the STA Satellite Max to the
LIS).

 Other codes are ignored by the STA Satellite
Max

1 ASCII character
‘X’: Order Cancelled (U)
‘O’: Order for Analysis (D)
‘F’: Final Results
‘Y’: No order on record (D)
‘Z’: No record for this patient
‘Q’: Response to a Query


Other codes are never sent by the STA Satellite
Max
Other codes are ignored by the STA Satellite
Max

Result Record
The Result Record is used by the STA Satellite Max to transmit assay results associated with a given sample
(Order) to the LIS.
CLSI LIS2-A2
(ASTM 1394)
Value
Field Description
Format
U
D
Comment
Section
9.1 (10.1.1)
‘R’
Record Type ID
1 ASCII character
 
9.2 (10.1.2)
>1
Sequence Number
N ASCII characters
  Field size not limited
12/21
0932500_EN
STA Satellite Max
CLSI LIS2-A2
(ASTM 1394)
ASTM Protocol
Value
Field Description
Format
U
D
Comment
Section
Universal Test ID
9.3 (10.1.3)
^
^
^
Test Code (Text)
^
‘#’,’F’
^
‘48’ or ‘64’
^
‘90’, ‘91’ or ‘92’
Assay Code
5 ASCII characters max
Result Type
1 ASCII character
‘F’: Final Result
‘#’: ‘1’ or ‘2’ for duplicates
Plasma Type
Ignored
^
The value of the Test Code
(component 4) is user defined in
the Assay Definition. (See 7.2)
STA Satellite Max Universal
Test ID max length is 5
‘48’: Control
‘64’: Specimen

Component Name
2 ASCII characters
‘90’ : Control Level 1
‘91’ : Control Level 2
‘92’ : Control Level 3
Dilution
‘X:Y’ NULL if no Dilution
^
Dilution (Text)
^

Component 1-3 are
NULL
Reagent Lot Number
N/A
Plasma Lot Number
9.4 (10.1.4)
Ignored
N/A
Value (Text)
Data or
255 ASCII characters max
Measurement Value

When a Result is not available
due to an error, this field contains
the text :
'NO RESULT'
(Multiple results are supported)
9.5 (10.1.5)
Unit (Open Text)
9.6 (10.1.6)
Units
255 ASCII characters max

‘|’
Reference ranges
/

Not used
9.7 (10.1.7)
‘|’
Result Abnormal
Flags
/

Not used
9.8 (10.1.8)
‘|’
Nature of
abnormality testing
/

Not used
9.9 (10.1.9)
'F’
Result Status
1 ASCII character, multiple status
separated by commas

Only ‘Final Result’ supported
‘F’: Final result
9.10 (10.1.10)
‘|’

/

Not used
Operator
Identification
255 ASCII characters

This field will contain the user
name under which the STA
Satellite Max session is open at
the result is sent
Not used
Date of Change in
Instrument
Normative Values
or Units
9.11 (10.1.11)
UserName (Text)
9.12 (10.1.12)
‘|’
Date/Time test
started
/

9.13 (10.1.13)
Date-Time
Date/Time Test
Completed
ASCII ‘YYYYMMDDHHMMSS’

0932500_EN
xxxx
13/21
20
STA Satellite Max
8.5
ASTM Protocol
Comment Record
The Comment Records contains
CLSI LIS2-A2
(ASTM 1394)
Format
U
D
Value
Field Description
Comment
10.1 (11.1.1)
‘C’
Record Type ID
1 ASCII character
 
10.2 (11.1.2)
>1
Sequence
Number
N ASCII characters
  Field size not limited
Section
10.4 (11.1.4)

Comment Text
(Text)
Comment Code
255 ASCII characters max

Comment Text
255 ASCII characters max

In download mode, the Comment Text field is
ignored by the STA Satellite Max
^
(Text)
* In Upload mode, depending on the Comment
Code, the Comment Text Component is either
clear text describing the Comment Code or the
internal error code that is displayed in the result
detail in the GUI *
10.5 (11.1.5)
‘I’, ‘G’
‘G’
Comment Type
1 ASCII character
1 ASCII character


In download mode comments are only accepted
for Order Records
‘I’: Flag
‘G’: Generic
* [STX]4R|1|^^^2^F^64|2.67|g/l||||F||Manufacturer||20180215115247[CR][ETX]B5[CR]
[STX]5C|1||-4001^OWREN-KOLLER buffer manally identified|I[CR][ETX]46[CR]
[STX]6C|2||-4003^STA-FIB 5 manually identified|I[CR][ETX]04[CR]
[STX]7C|3||-3004^Test Co 08 Result > is greater than verification high value|I[CR][ETX]FB[CR]
[STX]0C|4||36^Test Co 08 controls are invalid|I[CR][ETX]EE[CR]
Table of Errors and Flag Codes: Please refer to section 8.10
14/21
0932500_EN
STA Satellite Max
8.6
ASTM Protocol
Request Information Record
The Query Information Record is used


By the STA Satellite Max : to query the Worklist associated with a given sample ID (Barcode ID) from
the LIS
By the LIS to query results associated with a given Sample (Order) from the STA Satellite Max
CLSI LIS2-A2
(ASTM 1394)
Format
U D
Value
Field Description
Comment
11.1 (12.1.1)
‘Q’
Record Type ID
1 ASCII character
 
11.2 (12.1.2)
>1
Sequence
Number
N ASCII characters
  Field size not limited
Section
11.3 (12.1.3)
Starting Range
ID Number
Ignored / N.T
'ALL' (text) can be used

Hospital ID
N/A
Barcode ID
255 ASCII characters
Ending Range
ID Number
N/A

^
(Text)
11.4 (12.1.4)
Ignored / N.T



Universal Test
ID
11.5 (12.1.5)
^
^
^
This field is NULL in Upload mode
Component 1-3 are
NULL
The value of the Test Code (component 4) is user
defined in the Assay Definition. (See 7.2)
5 ASCII characters max
Test Code (Text)
^
‘#’,’F’
^
STA Satellite Max Universal Test ID max length
is 5
Assay Code
Result Type
1 ASCII char
‘F’: Final Result
‘#’: ‘1’ or ‘2’ for duplicates
‘48’: Control
‘64’: Specimen
‘48’ or ‘64’
^
‘90’, ‘91’ or ‘92’
Plasma Type
Component Name
^
'ALL' (text) can be used

2 ASCII characters
‘90’ : Control Level 1
‘91’ : Control Level 2
‘92’ : Control Level 3
‘X:Y’ NULL if no Dilution
Dilution (Text)
^
Dilution
N/A
Ignored
^
Reagent Lot
Number
Ignored
Plasma Lot Number
N/A
11.6 (12.1.6)
‘R’
Ignored
Nature of
Request Time
Limits
1 ASCII character

1 ASCII character
Specifies that the Date/ Time specified in 11.7
 (12.1.7) and 11.8 (12.1.8) are Result Test
Date/Time
11.7 (12.1.7)
Date-Time
Beginning
ASCII ‘YYYYMMDDHHMMSS’ 
Request Results
date and Time

11.8 (12.1.8)
Date-Time
Ending Request ASCII ‘YYYYMMDDHHMMSS’ 
Results date
and Time

0932500_EN
xxxx
15/21
20
STA Satellite Max
CLSI LIS2-A2
(ASTM 1394)
Value
ASTM Protocol
Field Description
Format
U D
Comment
Section
‘N’, ‘O’, ‘D’, ‘A’
‘N’, ‘F’
11.13
(12.1.13)
Request
Information
Status Code
1 ASCII character


1 ASCII character
‘N’: Requesting New or Edited
result only
‘O’: Requesting Test Orders
and Demographics only
‘D’: Requesting Demographics
Only
‘F’: Final Results
‘A’: Abort/Cancel Last
Request
8.7
Message Terminator Record
The Terminator Record Terminates a message. It is mandatory.
CLSI LIS2-A2
(ASTM 1394)
Format
U D
Value
Field Description
Comment
12.1 (13.1.1)
‘L’
Record Type ID
1 ASCII character
 
12.2 (13.1.2)
‘1’
Sequence
Number
1 ASCII characters
 
12.3 (13.1.3)
‘N’
Termination
Code
1 ASCII characters
  Specifies a Normal Termination
Section
16/21
0932500_EN
STA Satellite Max
8.8
ASTM Protocol
Manufacturer Information Record
The Manufacturer Information record contains the lot numbers of the products that have been used
to perform an assay.
The Manufacturer Record is transmitted by the Satellite Max. It is always associated with a Result
Record and follows immediately after the Result Record and after the associated Comments
Records if any.
Only one Manufacturer record can be sent with a given result.
If a Manufacturer Record is received by the STA Satellite Max it will be ignored.
CLSI LIS2-A2
(ASTM 1394)
Value
Field Description
Format
U
D
Comment
Section
14.1 (15.1.1)
‘M’
Record Type ID
1 ASCII character

14.2 (15.1.2)
>1
Sequence
Number
N ASCII characters

Field size not limited
5 ASCII characters max

The value of the Test Code (component 4) is
user defined in the Assay Definition.
255 ASCII characters max

Test Code (Text)
Product name
^
Assay Code
Diluent
20 ASCII characters max
Lot Number
Product name
^
Decontaminant
Reagent 1
Reagent 2
Reagent 3
Start Reagent
Lot Number

255 ASCII characters max

255 ASCII characters max

20 ASCII characters max
Lot Number
Product name
^
255 ASCII characters max
All fields, including unused fields should be
transmitted.
Unused fields should be left blank.
20 ASCII characters max
Lot Number
Product name
^

20 ASCII characters max
Lot Number
Product name
^
255 ASCII characters max
20 ASCII characters max
Lot Number
Product name
^

20 ASCII characters max
Lot Number
Product name
^
255 ASCII characters max
Control
255 ASCII characters max
20 ASCII characters max
0932500_EN
xxxx

If the Manufacturer Record is associated with
a control Result, the Control lot Number
should be placed in this field, otherwise the
field should be left blank
17/21
20
STA Satellite Max
8.9
ASTM Protocol
Error and Flag Code Table
Code
Comment
Type
Result
STA Satellite Max Common codes
36
Invalid Assay controls
Alarm
Result transmitted
-691
Un-calibrated assay
Alarm
Result transmitted
-2232
No Clot Error
Alarm
Result transmitted
STA Satellite Max Specific Errors Codes
-3001
Panic Low Result Flag
Alarm
Result transmitted
-3002
Panic High Result Flag
Alarm
Result transmitted
-3003
Low Result Flag
Alarm
Result transmitted
-3004
High Result Flag
Alarm
Result transmitted
-3005
Duplicate Error
Error
Mean value not transmitted*.
Results 1 and 2 transmitted.
-3006
Linearity Error
Error
Result not transmitted*
-3007
Calculation error
Error
Result not transmitted*
-3008
Calibration Range
Error
Result not transmitted*
-3009
Parallelism Error
Error
Result not transmitted*
-3010
Missing product Error
Error
Result not transmitted*
-3011
Optical Measurement Error (Noise)
Error
Result not transmitted*
-3012
No Result Found Error
Error
Result not transmitted*
-3013
Positive Hook effect
Error
Result not transmitted*
-3014
Optical Interferent in reaction
Alarm
Result transmitted
-3015
No level detection
Alarm
Result transmitted
-3016
Stirring
Alarm
Result transmitted
-3017
Initial Tray
Error
Result not transmitted*
-3018
Final Tray
Error
Result not transmitted*
-3019
Jump
Error
Result not transmitted*
-4001
Product / Buffer Identified Manually
Alarm
Result transmitted
-4002
Product / Cleaner Identified Manually
Alarm
Result transmitted
-4003
Product / Starter Identified Manually
Alarm
Result transmitted
-4004
Product / Reagent 1 Identified Manually
Alarm
Result transmitted
-4005
Product / Reagent 2 Identified Manually
Alarm
Result transmitted
-4006
Product / Reagent 3 Identified Manually
Alarm
Result transmitted
-4007
Sample identified manually
Alarm
Result transmitted
5002
Optical Reader Error
Error
Result not transmitted*
5003
Other Module Error**
Error
Result not transmitted*
* When a Result is not available due to an error, this field contains the text: 'NO RESULT'
** Code 5003: Module problem or insufficient quantity of plasma (QNS), see the screen analyzer for more accuracy.
18/21
0932500_EN
STA Satellite Max
ASTM Protocol
9 Examples of Messages sent to the LIS
In the following, the example frames are given in full, including the framing characters and checksum.
For a complete specification of the Low level Communication Specification please refer to the LIS01A2 Document.
9.1
Queries
9.1.1 Worklist Query
The STA Satellite Max sends this message to the LIS if the option ‘Host Query Enabled’ is activated and in the
following cases:
After a bar-coded sample is loaded.
After an unidentified sample has been given an ID manually.
Example: Query Work Load for the sample with Barcode ID = 100250
[Stx]1H|\^&|||STA_Satlellite_Max^08M10110 ^1|||||||P|1|20090701105221[CR][Etx]FF[CR]
[Stx]2Q|1|^100250||ALL||||||||O[CR][Etx]44[CR]
[Stx]3L|1|N[CR][Etx]00[CR]
9.1.2 Worklist Query Cancelled
If No Worklist has been received for this sample after the time-out defined in the LIS setup, the STA Satellite
Max sends a cancel
query message:
[Stx]1H|\^&|||STA_Satlellite_Max^08M10110 ^1|||||||P|1|20090701105234[CR][Etx]FF[CR]
[Stx]2Q|1|^100250||ALL||||||||A[CR][Etx]33[CR]
[Stx]3L|1|N[CR][Etx]00[CR]
9.2
Results Upload
9.2.1 Patient Result Upload
No Patient Information
Sample Barcode: 300105
In Carousel: 0902-0901
At Position: 3
Sequence Number: 205
Assay LIS Code: 10011
Result: 17.50
Unit: Second
User Name: UserName
Reagent Used: STA-TROMBIN 10 (Starter)
Reagent Lot: 250076
[Stx]1H|\^&|||STA_Satlellite_Max^0007M10052^1|||||||P|1|20090504141209[CR][Etx]1E[CR]
[Stx]2P|1[CR][Etx]3F[CR]
[Stx]3O|1|300105^^0902-0901^3|205||R||||||||||||||||||||F[CR][Etx]46[CR]
[Stx]4R|1|^^^10011^F^64|17.50|S||||F||UserName||20090504141208[CR][Etx]88[CR]
[Stx]5M|1|10011||||||STA-TROMBIN 10^250076[CR][Etx]F5[CR]
[Stx]6L|1|N[CR][Etx]09[CR]
0932500_EN
xxxx
19/21
20
STA Satellite Max
ASTM Protocol
9.2.2 Patient Result Upload with Alarms and Patient Information
Patient Last Name: Smith
Patient First Name: John
Patient Birth Date: 24 January 1985
Physician Name: Steve Henderson
Patient Location: 1258
Sample Barcode: 302157
In Carousel: 0902-0901
At Position: 6
Sequence Number: 210
Assay LIS Code: 10011
Result: 17.50
Unit: Second
User Name: UserName
Reagent Used: STA-TROMBIN 10 (Starter)
Reagent Lot: 250076
[Stx]1H|\^&|||STA_Satlellite_Max^0007M10052^1|||||||P|1|20090504141209[CR][Etx]1E[CR]
[[Stx]2P|1||||Smith^John||19850124||||||Steve Henderson||||||||||||1258[CR][Etx]0C[CR]
[Stx]3C|1||-4007^Sample has been manually identified |I[CR][Etx]96[CR]
[Stx]4O|1|302157^^0902-0901^6|210||R||||||||||||||||||||F[CR][Etx]4F[CR]
[Stx]5R|1|^^^10011^F^64|13.20|S||||F||UserName||20090504141230[CR][Etx]7D [CR]
[Stx]6C|1||36^Assay ATT10 s controls are invalid|I[CR][Etx]E5[CR]
[Stx]7M|1|10011||||||STA-TROMBIN 10^250076[CR][Etx]F7[CR]
[Stx]0L|1|N[CR][Etx]03[CR]
9.2.3 Patient Result Upload – Result in Duplicate
No Patient Information
Sample Barcode: 305521
In Carousel: 0902-0901
At Position: 7
Sequence Number: 212
Assay LIS Code: 10011
Result (Mean Value): 15.65
Unit: Second
User Name: UserName
Reagent Used: STA-TROMBIN 10 (Starter)
Reagent Lot: 250076
Result 1: 15.90 s
Result 2: 15.40 s
[Stx]1H|\^&||| STA_Satlellite_Max^0007M10052^1|||||||P|1|20090506103359[CR][Etx]24[CR]
[Stx]2P|1|[CR][Etx]BB[CR]
[Stx]3O|1|305521^^0902-0901^7|212||R||||||||||||||||||||F[CR][Etx]4F[CR]
[Stx]4R|1|^^^10011^F^64|15.65|S||||F||UserName||20090506103359[CR][Etx]93[CR]
[Stx]5M|1|10011||||||STA-TROMBIN 10^250076[CR][Etx]F5[CR]
[Stx]6R|2|^^^10011^1^64|15.90|S||||F||UserName||20090506103358[CR][Etx]7E[CR]
[Stx]7R|3|^^^10011^2^64|15.40|S||||F||UserName||20090506103359[CR][Etx]7D[CR]
[Stx]0L|1|N[CR][Etx]03[CR]
20/21
0932500_EN
STA Satellite Max
ASTM Protocol
9.2.4 Control Result Upload
Assay Internal Code: TQ_S
Control: Level 1
Sequence Number: 235
Assay LIS Code: 10011
Result: 13.50
Unit: Second
User Name: UserName
Reagent Used: STA-TROMBIN 10
Reagent Lot: 250076
Control Used: STA-COAG CONTROL N
Control Lot: 251070
[Stx]1H|\^&|||STA_Satlellite_Max^0007M10052^1|||||||P|1|20090504145243[CR][Etx]20[CR]
[Stx]2P|1||||TQ S^L_1[CR][Etx]81[CR]
[Stx]3O|1|TQ S L_1|235[CR][Etx]E5[CR]
[Stx]4R|1|^^^10011^F^48^90|13.50|S||||F||UserName||20090504145243[CR][Etx]50[CR]
[Stx]5M|1|10011|||||| STA-TROMBIN 10^250076|VERIFY CONTROL 1^S111557[CR][Etx]DD[CR]
[Stx]6L|1|N[CR][Etx]09[CR]
0932500_EN
xxxx