UNIVERSAL CENTRIFUGE Model: PLC-025 USER’S MANUAL OPERATION / MAINTENANCE INSTRUCTION & PARTS LIST ! Safety Information This manual contains important and safety information. The user must carefully read and understand the contents of this manual prior to using this equipment. 260101.1000 UNIVERSAL CENTRIFUGE Model: PLC-025 USER’S MANUAL OPERATION / MAINTENANCE INSTRUCTION & PARTS LIST CONTENT Safety information …………………………………………………… Alert instruction …………………………………………………… CAUTIONS …………………………………………………… WARNINGS …………………………………………………… Classification ………………………………………………….. . Standard …………………………………………………… Intended use …………………………………………………… Definition …………………………………………………… Content …………………………………………………… Specification …………………………………………………… Buckets …………………………………………………… Features ………………………………………………….. Installation ………………………………………………….. CAUTION ………………………………………………….. Control panel ………………………………………………….. Lid-Open Override ………………………………………………….. Rotor assembly ………………………………………………….. Buckets & Rotors ………………………………………………….. CAUTIONS ………………………………………………….. Preparation ………………………………………………….. Function Diagram ………………………………………………….. Operation ………………………………………………….. How to open LID ………………………………………………….. Maintenance & Servicing ………………………………………………….. Troubleshooting ………………………………………………….. ERRORS detection ………………………………………………….. Direction to Clean ………………………………………………….. Transportation ………………………………………………….. Storage …………………………………………………. Conditions for Transportation & Storage …………………………………….. Diagram of Components …………………………………………………. Parts List …………………………………………………. Wiring diagram …………………………………………………. One Year Limited Warranty …………………………………………………. 【ISO9001, ISO13485, GMP, FDA, CE marking Certified】 Gemmy Industrial Corp. Taipei, Taiwan 1 1 1 1 2 2 2 2 2 3 3 3 4 4 5 6 7 8 8 9 10 11 11 12 14 14 15 15 15 15 16 17 18 19 Safety Information Your GEMMY Universal Centrifuge has been designed with functions, reliability, and safety in mind. This manual contains important operation and safety information. The users must carefully read and understand the contents of this manual prior to installing and using this centrifuge in conformance with your local electrical codes. Always keep this manual along with the centrifuge. ALERT instruction WARNING Warning alerts you to a possibility of personal injury. CAUTION Cautions alert you to a possibility of damage to the equipment. NOTE Notes alert you to pertinent facts and condition. CAUTIONS To avoid electrical shock, always: 1、Use a properly grounded electrical outlet of correct voltage and current handling capacity. 2、Disconnect from power supply before maintenance and servicing. Refer to servicing to qualified personnel and engineers. 3、In the event that fluid is accidentally spilled into the chamber, disconnect grounded plug and turn the equipment upside down to avoid fluid contacting the internal components. WARNINGS To avoid personal injury: 1、Ensure that this equipment is connected to electrical service according to local and national electrical codes. Failure to properly connect may create a fire or shock hazard. 2、Always use the gloves and handle to remove the substances out of the chamber in order to protect from the possible virus infection. 3、Don’t open the centrifuge lid only until the motor is stopped completely. 1 Classification:This Centrifuge is classified as “Class I” equipment. CE marking:The manufacturer hereby declares the conformity with the actual Required safety standards in according with 93/42 EEC. Standard: This equipment complies with EC Electromagnetic Compatibility requirement of EN60601-1-2. EMI filter is designed in conformity with 89/336 EEC-EMC Directive. This equipment complies EN standard. ISO9001, ISO13485 GMP (Good Manufacturing Practice) Intended use: This centrifuge is to provide a desired testing within the range of 100 – 5000rpm for test tubes containing in vitro diagnostic specimens for blood separation for qualitative or quantitative test procedures. It is also applied to biochemistry, microbic immunity genetics and a general clinic test Definition: g Force the measurement for samples undergoing the stress of acceleration in a centrifuge RCF (Relative Centrifugal Force) RCF is the centrifugal force that the samples undergoes which relies on the speed of rotation ( N ) in rpm and the rotating radius ( R ). The rotating radius is measured from the center axis of the rotor to the extreme end of the centrifuge tubes. Formula: RCF = 11.18 x 4 x (N / 1000)² N:Speed of rotation R:Rotating radius in cm CONTENT: Your GEMMY centrifuge PLC-025 contains: - Detachable power code 1 - Rotor 1 - Bucket (Must be ordered separately) M-1560 : 60 tubes of 1.5ml capacity M-2060: 60 tubes of 2.0ml capacity M-0548: 48 tubes of 5.0ml capacity T-1016: 16 tubes of 10ml capacity T-1516: 16 tubes of 15ml capacity T-504 : 4 tubes of 50ml capacity (Other capacities of tubes are available upon request) 2 Specification PLC-025 360 x 450 x 292mm 20.4 kg (44.88 lbs) approx. 200 – 5000 rpm adjustable, LED display 4124 x g, LED display Swing out (horizontal) Brush-less, 300 W 2-digits, 99 min. LED display Yes Yes (6 errors detection) Automatic braking system Yes Yes, manual Yes Yes 110V or 220V, 50/60Hz ISO9001, ISO13485, CE, GMP Code Overall dimension New weight Speed range Max RCF Rotor type Motor Timer Imbalance detection Errors detection Brake Interlocking safety lid Lie open override device Safety shutout device Over heat alarm Power Quality Standard Buckets Bucket M-1560 M-2060 M-0548 T-1016 T-1516 T-5004 T-1004 (Ordering information) Capacity 1.5ml 2.0ml (or 2.2ml) 5.0ml 10ml 15ml 50ml 100ml Nr. Of tubes 60 tubes 60 tubes 48 tubes 16 tubes 16 tubes 4 tubes 4 tubes Max RPM 4000 4000 4000 5000 5000 5000 5000 Max RCF 2639 x g 2639 x g 2639 x g 4124 x g 4124 x g 4124 x g 4124 x g NOTE:Other volumes of tubes are available, please specify when ordering. 【Specifications are subject to change】 Features Swing-out rotor Brush-less motor, maintenance free Imbalance detection RPM / RCF conversion automatically Errors detection Electronic interlocking lid safety system Lid-open override system Fine adjustment for time, RPM and RCF Automatic braking system Alarm to shut off power automatically 3 Installation 1、 Remove the packing band from the carton box. 2、 Move the cover of the carton box firstly 3、 Move the center piece of the carton box 4、 Carry the centrifuge from the box and place onto a proper location. 5、 Move the plastic dust cover from the centrifuge 6、 Store and keep the cardboard packing box and reuse for transportation in future. Cover CAUTION: 1. Don’t drop the equipment, nor place the equipment upside down. 2. Do not place this equipment in the draft, sunlight or near a piece of equipment that emits heat and electromagnetic conduction. 3. Conditions for transportation & storage: Temperature 10ºC – 40ºC Humidity 40% - 90% 4. After unpacking the centrifuge, please check if the accessories are correct or not. 5. Check if there is any damage of the centrifuge during the transportation. If yes, contact your local dealer immediately. 6. Check the specification label located on back of this equipment, the power source must conform to your local codes. 4 Control panel Indication 1. LID LOCK : Indicate the proper lock of the lid 2. Alarm : Indicate imbalance rotor 3. SPIN : Indicate the rotor spinning. 4. SV : Press this pad indicating the set variable (SV) 5. RCF/RMP : Press this pad to monitor the value of RCF and RPM 6. STOP : Press this pad to stop the rotor spinning. 7. SV lamp : Indicate the display of set variable 8. RPM : Indicate the display of speed 9. RCF : Indicate the display of RCF 10. LID open : Press this pad to open the lid 11. START : When all values are set, press this pad to start spinning 12. ON/OFF : Power switch 13. TIME : Turn the controlled knob to adjust the desired time 14. SPEED : Turn the controlled knob to adjust the desired speed and RCF. NOTE:Always press key “LID OPEN” before opening the lid. 5 Lid-Open Override How to open the LID: Before unlocking the lid, please follow the procedures as stated below: 1、Switch on the power; 2、Press the key “LID OPEN” 3、Press the part # 15 of “Lid Release button”, then lift the lid up. In case of power failure, lid can’t be opened In case of failure of power supply, the lid locking system cannot unlock the lid automatically. On this occasion, the following procedures will help open the lid easily: 1、Switch off the power supply 2、Remove the plastic plug(#1)located on the left hand of the body. You will find a magnetic interlock fitting(#2)in it. 3、Insert a screwdriver(#3)or any steel bar into the hole 4、Press the magnetic interlock fitting down and also press the lock release button on the right hand(#4)at the same time. 5、Place the plastic plug(#1)back to close up the hole. WARNING: Disconnect from the power supply prior to maintenance. 6 Rotor assembly CAUTION: Inspect all components prior to installing this centrifuge 1. The locking bolt(#1) must be fastened firmly with Rotor + Drive shaft 2. The thread of locking bolt(#1)is designed by the reversed direction to the driving direction, thus never worry about the loose bolt. 3. Check if the pin is assembled firmly and properly in the drive shaft. 4. Rotate the rotor slightly to test the balance of the rotor. If the rotor is not locked firmly, the possible vibration and imbalance will occur, thus the damage or personal injury may be happened. 7 BUCKET and TUBES 【A】 The rubber cushion is used to protect the glass tubes during the high speed spinning. Thus make sure if the cushion is placed properly on the bottom of the bucket as depicted in figure 1. Let the concaved side of cushion up. Or clear all glass pieces if glass tube is broken. If the cushion is unfortunately placed upside down or on the tilt, recover it to the normal position. 【B】 The basket type bucket is also designed to place one rubber cushion in it. The purpose of this cushion is to avoid the plastic adaptor from vibration and crack. 【C】 How to place the centrifugal tubes CAUTION Use the same type of bucket for spinning in order to keep the balance of the rotor. Place the tubes diagonally across the bucket as stated in the figure Even if the less testing tubes is used, place all other tubes into the adaptor and buckets to allow the balance during spinning. 8 Preparation Before use of this centrifuge, the following steps must be conducted strictly: 1、Plug power cord in a grounded outlet. 2、Switch power to ON position. The power lamp ON/OFF and LID LOCK must be illuminated. 3、Press LID OPEN key then push Lid Release Button on the right hand of the housing to see if the lid can be opened smoothly or not. 4、Place the set bucket and tubes onto the rotor. NOTE: Different bucket or tubes in rotor is strictly prohibited. 5、Close the lid carefully when the LID LOCK lamp will be illuminated. If not, close the lid again. Otherwise, operation is impossible. 6、Adjust timer by turning the controlled knob to see if the timer can work or not. 7、Adjust the RPM by turning the controlled knob to see if the RPM can be functioned or not. 8、Press RPM/RCF key to monitor if RPM can be appeared or not, when if corresponding lamp RPM will be lit up. One more press on this key to monitor if RCF can be displayed or not when the corresponding lamp will be illuminated. 9、Press START key to see the rotor can spin or not when the corresponding lamp SPIN will be ON. 10、Press STOP key to see if the spinning can be stopped or not. 11、Press LID OPEN key and Lid release button to see if the lid can be opened or not. 12、Switch off the power If the above trial procedure is worked normally and smoothly, it indicates that this centrifuge is in good condition. You may operate this centrifuge anytime needed. 9 Function Diagram 1、Lock the lid 2、Press power switch (ON/OFF) 3、SV lamp is illuminated If not, open the lid and lock again 4、LID light lamped 5、Set time and set RPM / RCF The last one set values will appear on the screen. 6、Press START key(Functions START only after 10 sounds.) If cannot work starting, the Errors Detection will appear E-2. Corrective action: Check the error and start over again. 7、Timer starts counting.(count down) 8、Speed accelerates gently 9、SPIN light lamped 10、SV light OFF. 11、RPM / RCF light illuminated. 12、RPM / RCF light lamped If not, Error detection E-1 will appear. Corrective action: Press STOP key and check the cause. 13、Speed accelerates to the max If not, Error detection H-1 / H-3 / E-5 / E-6 will appear Corrective action: Press STOP key immediately and inspect SV value and press SV key again. 14、Time end:Timer and RPM/RCF appear 0 (ZERO) 15、3 sounds alert 16、SPIN light OFF 17、Press LID OPEN key If lid can’t be opened automatically, see Lid-Open Override in Page 6 18、Finish NOTE: NOTE: Operation may processedaccording accordingtotothe theabove abovesteps. Operation may bebe proceeded procedure. 10 OPERATION (Refer to Function Diagram, Page 9) 1、 Refer to the specification label located on back of this equipment. Be sure the power source must conform to your local codes. 2、 Plug cord in a properly grounded outlet. 3、 Switch power on and press LID OPEN key to open the lid. 4、 Choice of the intended bucket and tubes to meet your application purpose. 5、 Place the test tubes into the tube buckets. Be sure to place tubes diagonally to keep the rotor balanced. 6、 Close the lid completely. The LID LOCK light is illuminated. 7、 Set Time and RPM/RCF: Turn the controlled knob of TIME to your desired time. Turn the controlled knob of RPM/RCF to your desired values. (RPM and RCF are conversed automatically) NOTE: The set variables of the last operation will be appeared before your setting. Use the same time and RPM values if needed. 8、 Press START key to start spinning when the SPIN & RPM/RCF lamps will be illuminated, while the SV lamp will be extinguished. 9、 Press SV or RPM/RCF keys to monitor the set variables when the corresponding light will be lamped. 10、The program will detect the errors automatically and appear the signals on the screen: E-1:Rotor Imbalance. Alarm and shutout E-4:Connector to Motor disengaged E-5:Overload E-6:Lid opening during spinning or Disengaged connectors H-1:Driver over heat H-2:Motor over heat 11、The speed will accelerate to the max in 15 – 20 seconds. 12、When the set time has passed, the set values of TIME and RPM/RCF will be back to 0. 3 sounds will alert you the completion of the operation, then the SPIN lamp will be OFF automatically 13、Press LID OPEN key and Lid release button to open the lid. 14、Operation complete. NOTE: Refer to the description of Control panel、Rotor Assembly, Buckets & Tubes、 Lid-Open Override and Function Diagram as stated above prior to operation. How to open the Lid: Before unlocking the lid, please make sure to: 1、 Switch on the power; 2、 Press key LID OPEN 3、 Press the part # 15 of Lid Release Button and lift the lid up. 11 Maintenance & Servicing WARNING 1. To avoid electrical shock, always disconnect from power supply prior to maintenance & Servicing. 2. People other than the qualified personnel or licensed engineers are strictly prohibited from maintenance and servicing. 3. Always use a properly grounded electrical outlet of correct voltage and current handling capacity. 4. Inspect and find if there is any abnormal situation happened to the centrifuge, if yes, stop use of this unit and label “WARNING::Abnormal” on the centrifuge. Ask the qualified engineer to repair before it has become to the normal. DAILY a、Check the accessories: 1. Locking bolt: This part must be screwed completely. The loose bolt will result in vibration and imbalance which will cause the damage or personal injury. 2. Locking pin:The same way as locking bolt. 3. Bucket & Adaptor:Any crack or abrasion of these parts is strictly prohibited. b、 Check the lamps & LED screen: Switch on the power and see if all lamps and LED screen are in normal condition or not. Turn the controlled knob to make sure if TIME and RPM can be adjusted and functioned normally. c、Check the lid lock: Switch power to ON position, press LID OPEN key and door release button and see if the lid can be opened normally or not. d、 Check the visual appearance: Clean the exterior appearance before and after use of this centrifuge. Clean the rotor chamber if it is stained by the spilled by blood, urine and solution. Be sure to wear the latex gloves to clean the chamber. e、Check the hinges of lid: Check if the hinges are loose or not. Note, the loosen hinges may cause the abnormal operation or personal injury. f、 Sterilization If bucket, rotor or tube adaptor is contaminated by the blood, urine or solution, use steam sterilizer to autoclave all accessories. 12 MONTHLY Inspection categories: Rotor、Buckets、Adaptors、Programmable Function a. Rotor、 Buckets and Adaptors: Check if the above 3 parts have crack, abrasion, deformation or corrosion. If any of the above situation has happened, stop operation and replace the new one immediately. Continuity of use these defective parts may cause damage of the centrifuge or personal injury. b. Clean the above 3 parts: Loosen the bolt, nut and pin to remove the rotor from the drive shaft. Clean the related parts immersed in neutral detergent water, then clean with the distilled water or tap water, finally wipe it with cotton clothing and dry it naturally. Hot air or heater is prohibited for cleaning. c. Functions check: Switch on power, and check if the RPM and RCF can be conversed correctly and normally. Also check if timer can work and count down normally. YEARLY In addition to the maintenance of DAILY and MONTHLY, the following inspections must be done: a. Speed (RPM) Use a digital tachometer and check if the centrifuge is spinning at the right speed or not. b. Centrifugal Force:(RCF) Calculate if RCF in line with RPM is correct or not. stated in the manual to estimate the RCF. Use the formula c. Overall inspection Disassemble the centrifuge (limited to the qualified engineer or licensed personnel) carefully and check all components inside the centrifuge, such as motor, circuit board, fuse, screw, nut, grounding…. etc. CAUTION: The yearly maintenance and disassembly are limited to the qualified personnel or licensed engineers. 13 TROUBLESHOOTINGS Problem Can’t switch on Lid can’t be opened Possible cause Corrective action Wrong power source Check power supply Disconnect the power cord Plug in power cord Blown fuse Replace new one Disengagement of wiring Connect the wiring Burnout motor Replace motor Malfunction magnetic fitting Open lid as per Lid-Open automatically against the correct operation Override in this manual If power switch on Switch power on and Press key LID OPEN Imbalance rotor Vibration Balance the rotor use the same type of bucket in rotor. Speed can’t reach Loose locking bolt Tighten the bolt Overload or defective motor Check the motor Wrong power source Check the power supply Disconnect the power cord Plug in power cord Blown fuse Replace new one Disengagement of wiring Connect the wiring Burnout motor Replace new one Tubes didn’t place diagonally Place tubes correctly the maximum Motor can’t rotate Broken test tubes Erratic speed Didn’t place tubes all way down Place tube all way down Imbalance rotor Balance the rotor Lack of rubber cushion Insert rubber cushion Defective PCB Repair or replace new one ERRORS DETECTION: E-1:Imbalance E-2:Motor connection lost E-3:Driver over heat E-4:Over load E-5:Improper Lid lock 14 Direction to Clean: Disconnect from power supply prior to clean. Clean the equipment after use Use a moist cleaning cloth to clean the case, rotor and buckets Dry the equipment thoroughly before operation. Fire, heater or hot air is prohibited. Do not immerse this equipment in water Never use benzine or paint thinner for cleaning. Wear latex gloves for clean. Transportation: Fragile, Handle with Care Use no Hook Do not drop this equipment Do not place this equipment upside down Pack this equipment well for transportation. Prevent this equipment from vibrating. Keep it balanced during transportation. Storage: Do not place this equipment in the draft, sunlight or near a piece of equipment that emits electromagnetic conduction emission. Disconnect from power supply while store this equipment This equipment shall be stored under the condition of room temperature and humidity Do not place any heavy load on this equipment. Use plastic dust cover when store this equipment. Conditions for Transportation & Storage: Temperature 10ºC – 40ºC Humidity 40% - 90% NOTE: This equipment complies with EC electromagnetic compatibility requirement of EN60601-1-2. EMI filter is designed in conformity with 89/336 EEC-EMC Directive This equipment complies with IEC801 series regulations. 15 DIAGRAM OF COMPONENTS Model: PLC-025 16 PARTS LIST Code 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 Description Quantity Lid, upper, (SPCC #1.0) Hinges, lid (Bronze) Lid, inner, (SUS#304/1.0) Chamber, (SUS#304/1.0) Motor, (MCR-BL-2204D-1-A01) Rubber cushion Bottom mount, (PO#2.0) Rubber feet Micro switch, for balance purpose (VS10NO 31C2) Rack, micro switch (SPCC#1.0) Pin, (C type, 3*30) Silicon gasket, 85cm Back plate, (PO#2.0) AC rack + Fuse Lid release button (PO#2.0) PCB, motor driver Connector, (Driver + Control Board) Transformer, (220V/9V.18V) VR, timer (RV24YN.B104) VR, speed (RV24YN.B104) Circuit board (Control board) Control panel, inner (PO#1.5) Power switch (13R-135B) Knob, timer (PN-8E) Knob, speed (PN-8E) Control panel (PVC) Micro switch, Lid (VS10NO 31C2) Lock (left / Right) (PO#2.0) Spacer, Lock (Bronze) Magnetic (A8D724V/12W) Rack, lock (PO#2.0) Push plate, lock (PO#2.0) Plug, (Lid-Open override) Case, steel (PO#2.0) Rotor yoke, (SUS#304) Bolt, rotor + shaft (SUS# M10) Lock, Hook (SUS#304) Shield, motor (Rubber) 1 2 1 1 1 3 1 4 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 2 2 1 1 1 1 1 1 1 1 1 NOTE:When ordering for replacement, be sure to order by its parts number and description 17 WIRING DIAGRAM PLC-025 18 NEW VERSION ONE YEAR LIMITED WARRANTY Your GEMMY product is guaranteed be to free from defects in materials and workmanship for one ( 1 ) year under normal use from the date of purchase. This WARRANTY does not apply to any product damaged by accident, misuse, abuse, improper line voltage, drop, fire, flood or if the product was altered or repaired by anyone other than the qualified service personnel. The liability of Gemmy Industrial Corp. is limited to repair or replacement and under no circumstances shall GEMMY be liable for any collateral consequential damages or losses. This guarantee specifically excludes the expendables and consumables. All warranty claims must be directed to the distributors or agents authorized by Gemmy Industrial Corp. responsible for the sale of this equipment. The users are responsible for shipping expenses. ----------------------------------------------------------------------------------------------------User Name : Address : Fax No. : -------------------------------------------------------------------------------------------------------------User’s complaint: -------------------------------------------------------------------------------------------------------------Date of Purchase: Product Model : Serial No. : Distributor : --------------------------------------------------------------------------------------------------------Manufacturer: EU Representative: Gemmy Industrial Corp. 7Fl. #106, Min Chuan W. Road Taipei, Taiwan A LEE s.a. 23 rue de Rotterdam 4000 Liege, Belgium Tel: 32-4-2535989 Fax: 32-4-2536824 ISO9001 ISO9001 ISO13485 ISO13485 GMP GMP FDA FDA CE CE