UNIVERSAL CENTRIFUGE
Model: PLC-025
USER’S MANUAL
OPERATION / MAINTENANCE
INSTRUCTION & PARTS LIST
! Safety Information
This manual contains important and safety information. The user must carefully
read and understand the contents of this manual prior to using this equipment.
260101.1000
UNIVERSAL CENTRIFUGE
Model: PLC-025
USER’S MANUAL
OPERATION / MAINTENANCE
INSTRUCTION & PARTS LIST
CONTENT
Safety information
……………………………………………………
Alert instruction
……………………………………………………
CAUTIONS
……………………………………………………
WARNINGS
……………………………………………………
Classification
………………………………………………….. .
Standard
……………………………………………………
Intended use
……………………………………………………
Definition
……………………………………………………
Content
……………………………………………………
Specification
……………………………………………………
Buckets
……………………………………………………
Features
…………………………………………………..
Installation
…………………………………………………..
CAUTION
…………………………………………………..
Control panel
…………………………………………………..
Lid-Open Override
…………………………………………………..
Rotor assembly
…………………………………………………..
Buckets & Rotors
…………………………………………………..
CAUTIONS
…………………………………………………..
Preparation
…………………………………………………..
Function Diagram
…………………………………………………..
Operation
…………………………………………………..
How to open LID
…………………………………………………..
Maintenance & Servicing
…………………………………………………..
Troubleshooting
…………………………………………………..
ERRORS detection
…………………………………………………..
Direction to Clean
…………………………………………………..
Transportation
…………………………………………………..
Storage
………………………………………………….
Conditions for Transportation & Storage ……………………………………..
Diagram of Components
………………………………………………….
Parts List
………………………………………………….
Wiring diagram
………………………………………………….
One Year Limited Warranty ………………………………………………….
【ISO9001, ISO13485, GMP, FDA, CE marking Certified】
Gemmy Industrial Corp.
Taipei, Taiwan
1
1
1
1
2
2
2
2
2
3
3
3
4
4
5
6
7
8
8
9
10
11
11
12
14
14
15
15
15
15
16
17
18
19
Safety Information
Your GEMMY Universal Centrifuge has been designed with functions,
reliability, and safety in mind. This manual contains important operation and
safety information. The users must carefully read and understand the
contents of this manual prior to installing and using this centrifuge in
conformance with your local electrical codes. Always keep this manual
along with the centrifuge.
ALERT instruction
WARNING
Warning alerts you to a possibility of personal injury.
CAUTION
Cautions alert you to a possibility of damage to the equipment.
NOTE
Notes alert you to pertinent facts and condition.
CAUTIONS
To avoid electrical shock, always:
1、Use a properly grounded electrical outlet of correct voltage and
current handling capacity.
2、Disconnect from power supply before maintenance and
servicing. Refer to servicing to qualified personnel and
engineers.
3、In the event that fluid is accidentally spilled into the chamber,
disconnect grounded plug and turn the equipment upside down
to avoid fluid contacting the internal components.
WARNINGS
To avoid personal injury:
1、Ensure that this equipment is connected to electrical service
according to local and national electrical codes. Failure to properly
connect may create a fire or shock hazard.
2、Always use the gloves and handle to remove the substances out of
the chamber in order to protect from the possible virus infection.
3、Don’t open the centrifuge lid only until the motor is stopped
completely.
1
Classification:This Centrifuge is classified as “Class I” equipment.
CE marking:The manufacturer hereby declares the conformity with the
actual
Required safety standards in according with 93/42 EEC.
Standard:
 This equipment complies with EC Electromagnetic Compatibility
requirement of EN60601-1-2. EMI filter is designed in conformity with
89/336 EEC-EMC Directive. This equipment complies EN standard.
 ISO9001, ISO13485
 GMP (Good Manufacturing Practice)
Intended use:
This centrifuge is to provide a desired testing within the range of 100 – 5000rpm
for test tubes containing in vitro diagnostic specimens for blood separation for
qualitative or quantitative test procedures. It is also applied to biochemistry,
microbic immunity genetics and a general clinic test
Definition:
g Force
the measurement for samples undergoing the stress of acceleration in a
centrifuge
RCF (Relative Centrifugal Force)
RCF is the centrifugal force that the samples undergoes which relies on
the speed of rotation ( N ) in rpm and the rotating radius ( R ). The
rotating radius is measured from the center axis of the rotor to the
extreme end of the centrifuge tubes.
Formula:
RCF = 11.18 x 4 x (N / 1000)²
N:Speed of rotation
R:Rotating radius in cm
CONTENT:
Your GEMMY centrifuge PLC-025 contains:
- Detachable power code 1
- Rotor
1
- Bucket (Must be ordered separately)
M-1560 : 60 tubes of 1.5ml capacity
M-2060: 60 tubes of 2.0ml capacity
M-0548: 48 tubes of 5.0ml capacity
T-1016: 16 tubes of 10ml capacity
T-1516: 16 tubes of 15ml capacity
T-504 : 4 tubes of 50ml capacity
(Other capacities of tubes are available upon request)
2
Specification
PLC-025
360 x 450 x 292mm
20.4 kg (44.88 lbs) approx.
200 – 5000 rpm adjustable, LED display
4124 x g, LED display
Swing out (horizontal)
Brush-less, 300 W
2-digits, 99 min. LED display
Yes
Yes (6 errors detection)
Automatic braking system
Yes
Yes, manual
Yes
Yes
110V or 220V, 50/60Hz
ISO9001, ISO13485, CE, GMP
Code
Overall dimension
New weight
Speed range
Max RCF
Rotor type
Motor
Timer
Imbalance detection
Errors detection
Brake
Interlocking safety lid
Lie open override device
Safety shutout device
Over heat alarm
Power
Quality Standard
Buckets
Bucket
M-1560
M-2060
M-0548
T-1016
T-1516
T-5004
T-1004
(Ordering information)
Capacity
1.5ml
2.0ml (or 2.2ml)
5.0ml
10ml
15ml
50ml
100ml
Nr. Of tubes
60 tubes
60 tubes
48 tubes
16 tubes
16 tubes
4 tubes
4 tubes
Max RPM
4000
4000
4000
5000
5000
5000
5000
Max RCF
2639 x g
2639 x g
2639 x g
4124 x g
4124 x g
4124 x g
4124 x g
NOTE:Other volumes of tubes are available, please specify when ordering.
【Specifications are subject to change】
Features
 Swing-out rotor
 Brush-less motor, maintenance free
 Imbalance detection
 RPM / RCF conversion automatically
 Errors detection
 Electronic interlocking lid safety system
 Lid-open override system
 Fine adjustment for time, RPM and RCF
 Automatic braking system
 Alarm to shut off power automatically
3
Installation
1、 Remove the packing band from the carton box.
2、 Move the cover of the carton box firstly
3、 Move the center piece of the carton box
4、 Carry the centrifuge from the box and place onto a proper location.
5、 Move the plastic dust cover from the centrifuge
6、 Store and keep the cardboard packing box and reuse for
transportation in future.
Cover
CAUTION:
1. Don’t drop the equipment, nor place the equipment upside down.
2. Do not place this equipment in the draft, sunlight or near a piece of
equipment that emits heat and electromagnetic conduction.
3. Conditions for transportation & storage:
Temperature
10ºC – 40ºC
Humidity
40% - 90%
4. After unpacking the centrifuge, please check if the accessories are
correct or not.
5. Check if there is any damage of the centrifuge during the
transportation. If yes, contact your local dealer immediately.
6. Check the specification label located on back of this equipment, the
power source must conform to your local codes.
4
Control panel
Indication
1. LID LOCK : Indicate the proper lock of the lid
2. Alarm
: Indicate imbalance rotor
3. SPIN
: Indicate the rotor spinning.
4. SV
: Press this pad indicating the set variable (SV)
5. RCF/RMP : Press this pad to monitor the value of RCF and RPM
6. STOP
: Press this pad to stop the rotor spinning.
7. SV lamp : Indicate the display of set variable
8. RPM
: Indicate the display of speed
9. RCF
: Indicate the display of RCF
10. LID open : Press this pad to open the lid
11. START : When all values are set, press this pad to start spinning
12. ON/OFF : Power switch
13. TIME
: Turn the controlled knob to adjust the desired time
14. SPEED : Turn the controlled knob to adjust the desired speed and RCF.
NOTE:Always press key “LID OPEN” before opening the lid.
5
Lid-Open Override
How to open the LID:
Before unlocking the lid, please follow the procedures as stated below:
1、Switch on the power;
2、Press the key “LID OPEN”
3、Press the part # 15 of “Lid Release button”, then lift the lid up.
In case of power failure, lid can’t be opened
In case of failure of power supply, the lid locking system cannot unlock the lid
automatically. On this occasion, the following procedures will help open the
lid easily:
1、Switch off the power supply
2、Remove the plastic plug(#1)located on the left hand of the body.
You will find a magnetic interlock fitting(#2)in it.
3、Insert a screwdriver(#3)or any steel bar into the hole
4、Press the magnetic interlock fitting down and also press the lock release
button on the right hand(#4)at the same time.
5、Place the plastic plug(#1)back to close up the hole.
WARNING:
Disconnect from the power supply prior to maintenance.
6
Rotor assembly
CAUTION:
Inspect all components prior to installing this
centrifuge
1. The locking bolt(#1) must be fastened firmly with Rotor + Drive
shaft
2. The thread of locking bolt(#1)is designed by the reversed
direction to the driving direction, thus never worry about the
loose bolt.
3. Check if the pin is assembled firmly and properly in the drive
shaft.
4. Rotate the rotor slightly to test the balance of the rotor. If the
rotor is not locked firmly, the possible vibration and imbalance
will occur, thus the damage or personal injury may be happened.
7
BUCKET and TUBES
【A】
The rubber cushion is used to protect the
glass tubes during the high speed spinning.
Thus make sure if the cushion is placed
properly on the bottom of the bucket as
depicted in figure 1.
Let the concaved side of cushion up. Or
clear all glass pieces if glass tube is broken.
If the cushion is unfortunately placed upside
down or on the tilt, recover it to the normal
position.
【B】
The basket type bucket is also
designed to place one rubber cushion
in it. The purpose of this cushion is to
avoid the plastic adaptor from vibration
and crack.
【C】
How to place the centrifugal tubes
CAUTION
 Use the same type of bucket for spinning in
order to keep the balance of the rotor.
 Place the tubes diagonally across the
bucket as stated in the figure
 Even if the less testing tubes is used, place
all other tubes into the adaptor and buckets
to allow the balance during spinning.
8
Preparation
Before use of this centrifuge, the following steps must be conducted strictly:
1、Plug power cord in a grounded outlet.
2、Switch power to ON position.
The power lamp ON/OFF and LID LOCK must be illuminated.
3、Press LID OPEN key then push Lid Release Button on the right hand of
the housing to see if the lid can be opened smoothly or not.
4、Place the set bucket and tubes onto the rotor.
NOTE:
Different bucket or tubes in rotor is strictly prohibited.
5、Close the lid carefully when the LID LOCK lamp will be illuminated. If
not, close the lid again. Otherwise, operation is impossible.
6、Adjust timer by turning the controlled knob to see if the timer can work or not.
7、Adjust the RPM by turning the controlled knob to see if the RPM can be
functioned or not.
8、Press RPM/RCF key to monitor if RPM can be appeared or not, when if
corresponding lamp RPM will be lit up. One more press on this key to
monitor if RCF can be displayed or not when the corresponding lamp will
be illuminated.
9、Press START key to see the rotor can spin or not when the corresponding
lamp SPIN will be ON.
10、Press STOP key to see if the spinning can be stopped or not.
11、Press LID OPEN key and Lid release button to see if the lid can be
opened or not.
12、Switch off the power
If the above trial procedure is worked normally and smoothly, it indicates that
this centrifuge is in good condition. You may operate this centrifuge
anytime needed.
9
Function Diagram
1、Lock the lid
2、Press power switch (ON/OFF)
3、SV lamp is illuminated
If not, open the lid and lock again
4、LID light lamped
5、Set time and set RPM / RCF
The last one set values will appear on the screen.
6、Press START key(Functions START only after 10 sounds.)
If cannot work starting, the Errors Detection will appear E-2.
Corrective action: Check the error and start over again.
7、Timer starts counting.(count down)
8、Speed accelerates gently
9、SPIN light lamped
10、SV light OFF.
11、RPM / RCF light illuminated.
12、RPM / RCF light lamped
If not, Error detection E-1 will appear.
Corrective action: Press STOP key and check the cause.
13、Speed accelerates to the max
If not, Error detection H-1 / H-3 / E-5 / E-6 will appear
Corrective action: Press STOP key immediately and inspect SV value
and press SV key again.
14、Time end:Timer and RPM/RCF appear 0 (ZERO)
15、3 sounds alert
16、SPIN light OFF
17、Press LID OPEN key
If lid can’t be opened automatically, see Lid-Open Override in Page 6
18、Finish
NOTE:
NOTE:
Operation
may
processedaccording
accordingtotothe
theabove
abovesteps.
Operation
may
bebe
proceeded
procedure.
10
OPERATION
(Refer to Function Diagram, Page 9)
1、 Refer to the specification label located on back of this equipment. Be sure the
power source must conform to your local codes.
2、 Plug cord in a properly grounded outlet.
3、 Switch power on and press LID OPEN key to open the lid.
4、 Choice of the intended bucket and tubes to meet your application purpose.
5、 Place the test tubes into the tube buckets. Be sure to place tubes diagonally to
keep the rotor balanced.
6、 Close the lid completely. The LID LOCK light is illuminated.
7、 Set Time and RPM/RCF:
Turn the controlled knob of TIME to your desired time.
Turn the controlled knob of RPM/RCF to your desired values. (RPM and RCF
are conversed automatically)
NOTE:
The set variables of the last operation will be appeared before your setting.
Use the same time and RPM values if needed.
8、 Press START key to start spinning when the SPIN & RPM/RCF lamps will be
illuminated, while the SV lamp will be extinguished.
9、 Press SV or RPM/RCF keys to monitor the set variables when the
corresponding light will be lamped.
10、The program will detect the errors automatically and appear the signals on the
screen:
E-1:Rotor Imbalance. Alarm and shutout
E-4:Connector to Motor disengaged
E-5:Overload
E-6:Lid opening during spinning or Disengaged connectors
H-1:Driver over heat
H-2:Motor over heat
11、The speed will accelerate to the max in 15 – 20 seconds.
12、When the set time has passed, the set values of TIME and RPM/RCF will be
back to 0. 3 sounds will alert you the completion of the operation, then the
SPIN lamp will be OFF automatically
13、Press LID OPEN key and Lid release button to open the lid.
14、Operation complete.
NOTE:
Refer to the description of Control panel、Rotor Assembly, Buckets & Tubes、
Lid-Open Override and Function Diagram as stated above prior to operation.
How to open the Lid:
Before unlocking the lid, please make sure to:
1、 Switch on the power;
2、 Press key LID OPEN
3、 Press the part # 15 of Lid Release Button and lift the lid up.
11
Maintenance & Servicing
WARNING
1. To avoid electrical shock, always disconnect from power supply prior to
maintenance & Servicing.
2. People other than the qualified personnel or licensed engineers are strictly
prohibited from maintenance and servicing.
3. Always use a properly grounded electrical outlet of correct voltage and
current handling capacity.
4. Inspect and find if there is any abnormal situation happened to the
centrifuge, if yes, stop use of this unit and label “WARNING::Abnormal”
on the centrifuge. Ask the qualified engineer to repair before it has
become to the normal.
DAILY
a、Check the accessories:
1. Locking bolt: This part must be screwed completely. The loose bolt
will result in vibration and imbalance which will cause the
damage or personal injury.
2. Locking pin:The same way as locking bolt.
3. Bucket & Adaptor:Any crack or abrasion of these parts is strictly prohibited.
b、 Check the lamps & LED screen:
Switch on the power and see if all lamps and LED screen are in normal
condition or not.
Turn the controlled knob to make sure if TIME and RPM can be adjusted
and functioned normally.
c、Check the lid lock:
Switch power to ON position, press LID OPEN key and door release
button and see if the lid can be opened normally or not.
d、 Check the visual appearance:
Clean the exterior appearance before and after use of this centrifuge.
Clean the rotor chamber if it is stained by the spilled by blood, urine and
solution. Be sure to wear the latex gloves to clean the chamber.
e、Check the hinges of lid:
Check if the hinges are loose or not. Note, the loosen hinges may
cause the abnormal operation or personal injury.
f、 Sterilization
If bucket, rotor or tube adaptor is contaminated by the blood, urine or
solution, use steam sterilizer to autoclave all accessories.
12
MONTHLY
Inspection categories:
Rotor、Buckets、Adaptors、Programmable Function
a. Rotor、 Buckets and Adaptors:
Check if the above 3 parts have crack, abrasion, deformation or corrosion.
If any of the above situation has happened, stop operation and replace the
new one immediately. Continuity of use these defective parts may cause
damage of the centrifuge or personal injury.
b. Clean the above 3 parts:
Loosen the bolt, nut and pin to remove the rotor from the drive shaft.
Clean the related parts immersed in neutral detergent water, then clean
with the distilled water or tap water, finally wipe it with cotton clothing and
dry it naturally. Hot air or heater is prohibited for cleaning.
c. Functions check:
Switch on power, and check if the RPM and RCF can be conversed
correctly and normally. Also check if timer can work and count down
normally.
YEARLY
In addition to the maintenance of DAILY and MONTHLY, the following
inspections must be done:
a. Speed (RPM)
Use a digital tachometer and check if the centrifuge is spinning at the right
speed or not.
b. Centrifugal Force:(RCF)
Calculate if RCF in line with RPM is correct or not.
stated in the manual to estimate the RCF.
Use the formula
c. Overall inspection
Disassemble the centrifuge (limited to the qualified engineer or licensed
personnel) carefully and check all components inside the centrifuge, such
as motor, circuit board, fuse, screw, nut, grounding…. etc.
CAUTION:
The yearly maintenance and disassembly are limited to the qualified
personnel or licensed engineers.
13
TROUBLESHOOTINGS
Problem
Can’t switch on
Lid can’t be opened
Possible cause
Corrective action
Wrong power source
Check power supply
Disconnect the power cord
Plug in power cord
Blown fuse
Replace new one
Disengagement of wiring
Connect the wiring
Burnout motor
Replace motor
Malfunction magnetic fitting
Open lid as per Lid-Open
automatically against
the correct operation
Override in this manual
If power switch on
Switch power on and
Press key LID OPEN
Imbalance rotor
Vibration
Balance the rotor
use the same type of
bucket in rotor.
Speed can’t reach
Loose locking bolt
Tighten the bolt
Overload or defective motor
Check the motor
Wrong power source
Check the power supply
Disconnect the power cord
Plug in power cord
Blown fuse
Replace new one
Disengagement of wiring
Connect the wiring
Burnout motor
Replace new one
Tubes didn’t place diagonally
Place tubes correctly
the maximum
Motor can’t rotate
Broken test tubes
Erratic speed
Didn’t place tubes all way down
Place tube all way down
Imbalance rotor
Balance the rotor
Lack of rubber cushion
Insert rubber cushion
Defective PCB
Repair or replace new one
ERRORS DETECTION:
E-1:Imbalance
E-2:Motor connection lost
E-3:Driver over heat
E-4:Over load
E-5:Improper Lid lock
14
Direction to Clean:
 Disconnect from power supply prior to clean.
 Clean the equipment after use
 Use a moist cleaning cloth to clean the case, rotor and buckets
 Dry
the equipment thoroughly before operation. Fire, heater or hot air is
prohibited.
 Do not immerse this equipment in water
 Never use benzine or paint thinner for cleaning.
 Wear latex gloves for clean.
Transportation:
 Fragile, Handle with Care
 Use no Hook
 Do not drop this equipment
 Do
not place this equipment upside down
 Pack this equipment well for transportation.
 Prevent this equipment from vibrating. Keep it balanced during
transportation.
Storage:
 Do not place this equipment in the draft, sunlight or near a piece of
equipment that emits electromagnetic conduction emission.
 Disconnect from power supply while store this equipment
 This equipment shall be stored under the condition of room temperature
and humidity
 Do not place any heavy load on this equipment.
 Use
plastic dust cover when store this equipment.
Conditions for Transportation & Storage:
 Temperature
10ºC – 40ºC
 Humidity
40% - 90%
NOTE:
 This
equipment complies with EC electromagnetic compatibility
requirement of EN60601-1-2.
 EMI filter is designed in conformity with 89/336 EEC-EMC Directive
 This
equipment complies with IEC801 series regulations.
15
DIAGRAM OF COMPONENTS
Model: PLC-025
16
PARTS LIST
Code
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
Description
Quantity
Lid, upper, (SPCC #1.0)
Hinges, lid (Bronze)
Lid, inner, (SUS#304/1.0)
Chamber, (SUS#304/1.0)
Motor, (MCR-BL-2204D-1-A01)
Rubber cushion
Bottom mount, (PO#2.0)
Rubber feet
Micro switch, for balance purpose (VS10NO 31C2)
Rack, micro switch (SPCC#1.0)
Pin, (C type, 3*30)
Silicon gasket, 85cm
Back plate, (PO#2.0)
AC rack + Fuse
Lid release button (PO#2.0)
PCB, motor driver
Connector, (Driver + Control Board)
Transformer, (220V/9V.18V)
VR, timer (RV24YN.B104)
VR, speed (RV24YN.B104)
Circuit board (Control board)
Control panel, inner (PO#1.5)
Power switch (13R-135B)
Knob, timer (PN-8E)
Knob, speed (PN-8E)
Control panel (PVC)
Micro switch, Lid (VS10NO 31C2)
Lock (left / Right) (PO#2.0)
Spacer, Lock (Bronze)
Magnetic (A8D724V/12W)
Rack, lock (PO#2.0)
Push plate, lock (PO#2.0)
Plug, (Lid-Open override)
Case, steel (PO#2.0)
Rotor yoke, (SUS#304)
Bolt, rotor + shaft (SUS# M10)
Lock, Hook (SUS#304)
Shield, motor (Rubber)
1
2
1
1
1
3
1
4
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
2
2
1
1
1
1
1
1
1
1
1
NOTE:When ordering for replacement, be sure to order by its parts number and description
17
WIRING DIAGRAM
PLC-025
18
NEW VERSION
ONE YEAR LIMITED WARRANTY
Your GEMMY product is guaranteed be to free from defects in materials and
workmanship for one ( 1 ) year under normal use from the date of purchase.
This WARRANTY does not apply to any product damaged by accident,
misuse, abuse, improper line voltage, drop, fire, flood or if the product was
altered or repaired by anyone other than the qualified service personnel.
The liability of Gemmy Industrial Corp. is limited to repair or replacement and
under no circumstances shall GEMMY be liable for any collateral consequential
damages or losses. This guarantee specifically excludes the expendables and
consumables.
All warranty claims must be directed to the distributors or agents authorized
by Gemmy Industrial Corp. responsible for the sale of this equipment. The
users are responsible for shipping expenses.
----------------------------------------------------------------------------------------------------User Name :
Address
:
Fax No.
:
-------------------------------------------------------------------------------------------------------------User’s complaint:
-------------------------------------------------------------------------------------------------------------Date of Purchase:
Product Model :
Serial No.
:
Distributor
:
--------------------------------------------------------------------------------------------------------Manufacturer:
EU Representative:
Gemmy Industrial Corp.
7Fl. #106, Min Chuan W. Road
Taipei, Taiwan
A LEE s.a.
23 rue de Rotterdam
4000 Liege, Belgium
Tel: 32-4-2535989
Fax: 32-4-2536824
ISO9001
ISO9001
ISO13485
ISO13485
GMP
GMP
FDA
FDA
CE
CE