SERVICE TOOL
INSTALATION AND USER MANUAL
Desktop application ver.: 1.0
Manual ver.: 1.0
Create date: 27.11.2022
by Enbio Technology. Sp. Z o.o.
Requirments:
• operating system: Windows 7, 8,10 & 11
• connection interface: port USB 2.0
• display settings for best appeareance (control panel ->
display settings):
• USB cable 2.0 type A 2x male
INSTALLATION
1. Download EnbioServiceTool.zip
2. Unzip to your computer
3. Install driver coresponding to your operating system:
Operating system
File
Windows 7 32 bits
VCP_V1.5.0_Setup_W7_x86_32bits
Windows 7 64 bits
VCP_V1.5.0_Setup_W7_x64_64bits
Windows 8/10/11 32 bits
VCP_V1.5.0_Setup_W8_x86_32bits
Windows 8/10/11 64 bits
VCP_V1.5.0_Setup_W8_x64_64bits
4. Double click on Enbio Service Tool Installer.exe
5. After installation you can find shortcut on your
desktop.
6. Application is ready to use.
CONNECTION PROCEDURE
1. Turn off device (Enbio S / Pro
autoclave).
2. Connect device to your computer by
usb cable.
3. Power on device.
4. Click Find Device.
HINT 1
To check if driver is proper installed and
device is recognized by your computer
open Control Panel -> Device Manager.
HINT 2
Some operating system cannot recognized
device if bluetooth is turned on. Switch it
off and try again.
Counters tab - READING INFORMATION
Counters:
• Total process
• HEPA
• Service
• First process
Serial Number
Date & time on device
DIP Switch state
Firmware version
Click Refresh to read
current values
Counters tab – CHANGING VALUES
Clear log history
- Clear error history
(visible in service
screen)
Change serial number: write
value to save and click Change
Reset Service
- Reseting days and
cycles to service
Date & time: click Update to
change time to time from your
computer
Clear all
- Writing 0 to Total
process / HEPA /
Service
Counters:
Write Change value
to save in device
then click Change.
Change All: save Change value (Total process / HEPA / Service) to device
Counters tab – Replacement device
The limit is automatically reset
when EnbioServiceTool connects
to the device.
To change the cycle limit for the
Replacement Device enter the
value and click Write.
Option only avaiable if connected
device is recognized as
Replacement Device.
Control tab – control output & read sensors
Click ON to open valve
Click OFF to close valve
Control heaters and
pumps – ON / OFF
Locking door – DO NOT OPEN
DURING STERILIZATION
PROCESS
Read live data from
temperatures and
pressure sensors
Log tab – read log and decode errors
Choose file: select log
to read
Show basic
information from
log:
- Serial number
- Firmware ver.
- Process nb.
- Process type
- Error number
Decode errors: convert code to list of errors
Save to file: save
decoded log to file
Open chart:
visualize values from
log on chart. Select
data to be printed
(pressure, process /
chamber / steam
generator temp.) –
see next slide
Zoom or save chart to image file
Other tab – send log to Enbio service
By application you can directly send uploaded log
to Enbio Service Department
Fill:
• Service name or service
technician first & last name
• Service address
• Phone number
• Email contact address
• Short description of problem or
failure
Click Send.